• BMC anesthesiology · Feb 2017

    Clinical Trial

    Evaluation of the optimal cuff volume and cuff pressure of the revised laryngeal tube "LTS-D" in surgical patients.

    • Marc Kriege, Christian Alflen, Johannes Eisel, Thomas Ott, Tim Piepho, and Ruediger R Noppens.
    • Department of Anesthesiology, University Medical Center of the Johannes Gutenberg-University Mainz, Mainz, Germany.
    • BMC Anesthesiol. 2017 Feb 2; 17 (1): 19.

    BackgroundRecent case reports have indicated significant cuff overinflation when using the standard filling volume based on the manufacturer's recommendations in older models of laryngeal tubes. The aim of this study was to determine the minimum cuff pressure needed to perform standardized ventilation without leakage in the new, revised model of the laryngeal tube "LTS-D".MethodsAfter ethical approval, LTS-D was placed for ventilation in 60 anesthetized patients. The cuff was inflated to the recommended volume (#3: 60 ml, #4: 80 ml, and #5: 90 ml). After evaluation of the initial cuff pressure (CP), the CP was lowered in 10 cmH2O steps until a minimal cuff pressure of 30 cmH2O was achieved. The absence of an audible leak was required for a step-by-step reduction in the CP. Evacuated cuff volume, success rate, and airway injuries were documented. Data were expressed as medians (interquartile ranges [IQRs]). The comparison of CPs and cuff volumes was performed using the Mann-Whitney test.ResultsAfter initial inflation, the CP ranged from 105 cmH2O [90-120; #5] to 120 cmH2O [110-120; #3]. Lowering the CP to 60 cmH2O resulted in a reduced cuff volume ranging from 47 ml [44-54; #3] to 77 ml [75-82; #5] compared to the initial inflation (p < 0.001). Leakage occurred more frequently when the CP was lowered to 40 cmH2O compared to the initial inflation (44/54 [81%]; p < 0.01). Using a CP between 50 cmH2O and 60 cmH2O, a leakage rate of 3/54 (5%) was observed, compared to a rate of 11/54 (21%) when using a CP lower than 50 cmH2O. The overall success rate was 90%, and airway injury occurred in 7% of patients (4/60).ConclusionWe found significant overinflation of the revised LTS-D using the recommended volume for initial cuff inflation. A CP of 60 cmH2O was found to be sufficient for ventilation in the majority of patients evaluated. Checking and adjusting the CP in laryngeal tubes is mandatory to avoid overinflation.Trial RegistrationClinicalTrials.gov NCT02300337 . Registered: 20 November 2014.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.