- Jeffrey M Haynes and Robert G Fishwick.
- Pulmonary Function Laboratory, Elliot Health System, Manchester, New Hampshire. jhaynes@elliot-hs.org.
- Respir Care. 2022 Apr 1; 67 (4): 428432428-432.
BackgroundBlood gas quality control (QC) is an essential and mandatory part of a laboratory's quality plan. The acceptable QC range should be 2 SD from the mean value. The use of assayed QC material does not negate the responsibility of the laboratory to calculate the mean and 2 SD ranges of QC measurements for verification. Verifying assayed QC ranges is a Clinical Laboratory Improvement Amendment (CLIA) requirement. This study shows the results of assayed QC mean and 2 SD range verification from a blood gas analyzer.MethodsQC data from a blood gas analyzer were compared to manufacturer-provided mean and ranges. The percent difference between the measured mean and the manufacturer-provided mean was calculated to assess agreement. The measured SD was used to determine how many SD the manufacturer-provided ranges were from the measured mean.ResultsThe largest difference in mean values was 2.27% > the manufacturer-provided mean. Forty-eight percent of all mean value comparisons showed a difference of 0%, and 71% were < 1%. The manufacturer-provided ranges were considerably wider than the measured 2 SD range, ranging from 2.4-75 SD. None of the manufacturer-provided ranges were deemed acceptable for clinical use.ConclusionsOur analysis validates the CLIA mandate and American Association for Respiratory Care Clinical Practice Guideline recommendation that laboratories must verify manufacturer-provided QC means and ranges and adjust QC means and ranges to match the performance of their blood gas analyzer.Copyright © 2022 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.