Observational StudyDual-scale single marker calibration for digital templating of total hip arthroplasty in standing radiographs: a prospective clinical study.
- Christian Ries, Christoph Tobias Baltin, Stefan Haneder, Peer Eysel, Martin Hellmich, and Christoph Kolja Boese.
- Division of Orthopaedics, Department of Trauma and Orthopaedic Surgery, University Medical Center Hamburg-Eppendorf, Martinistr. 52, 20246, Hamburg, Germany.
- Arch Orthop Trauma Surg. 2023 Apr 1; 143 (4): 181718241817-1824.
PurposeAccuracy of calibration of radiographs significantly influences the quality of digital templating for total hip arthroplasty (THA). The standard of care is calibration with external calibration markers (ECM). This method is associated with significant errors. Dual-scale single marker (DSSM) calibration methods may improve accuracy. The present prospective observational study is the first to analyze the application of a DSSM method in standing pelvis radiographs.Methods100 patients with unilateral THA underwent antero-posterior pelvis radiographs with ECM and DSSM. The hip components were used as reference calibration factor (internal calibration factor; ICM). Absolute differences of calibration factors for ECM and DSSM from ICM were calculated. Absolute relative deviations (ARD) were calculated. Subgroup analysis for sex and WHO BMI category was performed. Furthermore, patients reported subjective comfort for each marker using a 10-point scale and choosing the preferred marker.ResultsMaximum magnification factor differences from the ICM were 23.3% and 9.5% and mean absolute differences were 12.5% and 2.1% for the ECM and DSSM, respectively. ARD from ICM was significantly lower for DSSM compared to ECM (p < 0.001). Absolute differences increased with BMI category using ECM; calibration by DSSM was consistent in all subgroups. Patients preferred DSSM over ECM (n = 53) or were indifferent (n = 20). Comfort was rated significantly higher for DSSM versus ECM (p < 0.001).ConclusionDSSM method showed superior results in comparison to the ECM method for calibration of digital radiographs. DSSM could be used to improve digital templating in standing radiographs.© 2022. The Author(s).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.