How precise is PreSize Neurovascular? Accuracy evaluation of flow diverter deployed-length prediction.
- Tufail Patankar, Jeremy Madigan, Jonathan Downer, Hemant Sonwalkar, Peter Cowley, and Francesco Iori.
- 1Department of Neuroradiology, Leeds Teaching Hospital, Leeds.
- J. Neurosurg. 2022 Oct 1; 137 (4): 107210801072-1080.
ObjectiveThe use of flow-diverting stents has been increasingly important in intracranial aneurysm treatment. However, accurate sizing and landing zone prediction remain challenging. Inaccurate sizing can lead to suboptimal deployment, device waste, and complications. This study presents stent deployment length predictions offered in medical software (PreSize Neurovascular) that provides physicians with real-time planning support, allowing them to preoperatively "test" different devices in the patient's anatomy in a safe virtual environment. This study reports the software evaluation methodology and accuracy results when applied to real-world data from a wide range of cases and sources as a necessary step in demonstrating its reliability, prior to impact assessment in prospective clinical practice.MethodsImaging data from 138 consecutive stent cases using the Pipeline embolization device were collected from 5 interventional radiology centers in the United Kingdom and retrospectively analyzed. Prediction accuracy was calculated as the degree of agreement between stent deployed length measured intraoperatively and simulated in the software.ResultsThe software predicted the deployed stent length with a mean accuracy of 95.61% (95% confidence interval [CI] 94.87%-96.35%), the highest reported accuracy in clinical stent simulations to date. By discounting 4 outlier cases, in which events such as interactions with coils and severe push/pull maneuvers impacted deployed length to an extent the software was not able to simulate or predict, the mean accuracy further increases to 96.13% (95% CI 95.58%-96.69%). A wide discrepancy was observed between labeled and measured deployed stent length, in some cases by more than double, with no demonstrable correlation between device dimensions and deployment elongation. These findings illustrate the complexity of stent behavior and need for simulation-assisted sizing for optimal surgical planning.ConclusionsThe software predicts the deployed stent length with excellent accuracy and could provide physicians with real-time accurate device selection support.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.