• Critical care medicine · Dec 1993

    Comparative Study Clinical Trial

    Respiratory deadspace measurements in neonates during extracorporeal membrane oxygenation.

    • J H Arnold, J E Thompson, and P K Benjamin.
    • Department of Respiratory Care, Children's Hospital, Boston, MA 02115.
    • Crit. Care Med. 1993 Dec 1;21(12):1895-900.

    ObjectivesTo evaluate the bias and precision of a simple, bedside method of quantification of minute CO2 production (study 1) and then apply the technique to measure physiologic deadspace in a group of neonates undergoing extracorporeal membrane oxygenation (ECMO) (study 2).DesignA prospective, clinical study comparing a simple method of quantifying minute CO2 production with a criterion standard (study 1); a cohort study evaluating the utility of deadspace measurements in neonates undergoing extracorporeal membrane oxygenation (study 2).SettingTertiary care pediatric intensive care unit (ICU) in a university hospital.PatientsStudy 1: Thirteen neonates (weighing from 2.5 to 6.2 kg) were studied during mechanical ventilation being administered for a variety of respiratory diseases. Study 2: Fifteen neonates with respiratory failure were examined during the course of ECMO. INTERVENTIONS (METHODS): Study 1: The CO2 concentration of expired gas sampled at the exhaust port of the test ventilator was continuously measured and transformed to minute CO2 production by the following formula: (calculated minute CO2 production) = (tension of CO2 in exhaust gas) x 0.0013 x (ventilator pneumotachometer minute volume), where 0.0013 is the conversion factor to express gas tension as fractional volume at standard conditions. Minute CO2 production was measured independently with a previously validated bedside calorimeter and the calculated and measured values were compared. Study 2: The CO2 concentration of expired gas sampled at the exhaust port of the test ventilator was continuously measured and transformed to mixed expired CO2 by the following formula, which corrects for compressible volume measured by the ventilator pneumotachometer: mixed expired CO2 = (tension of CO2 in exhaust gas) x (ventilator pneumotachometer minute volume)/(minute volume at proximal airway). We then utilized the Bohr-Enghoff method to calculate the deadspace/tidal volume ratio: deadspace/tidal volume ratio = (PaCO2--mixed expired CO2 tension)/PaCO2.Measurements And Main ResultsStudy 1: Calculated minute CO2 production ranged between 11.8 and 38.9 mL/min. When compared with measured minute CO2 production, the bias and precision were -0.8 and 1.7 mL/min, respectively. The mean percent difference for calculated minute CO2 production was 3.7%. Study 2: Deadspace/tidal volume ratio was calculated during temporary separation from ECMO support as the patients demonstrated improvements in gas exchange and lung compliance. Fifty-two measurements were made in 15 patients, and 13 of 15 patients demonstrated a decrease in deadspace/tidal volume ratio during the course of ECMO. The mean decrease was 21% and the decrease was statistically significant.ConclusionsMinute CO2 production can be measured simply and accurately, using equipment readily available in most ICU settings. The same method can be utilized to calculate the deadspace/tidal volume ratio, which provides valuable information regarding the gas exchanging efficiency of the neonatal lung during ECMO.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…