Randomized Controlled Trial
Performance comparison of intraosseous devices and setups for infusion of whole blood in a cadaveric swine bone model.
- Katherine J Lee, Morgan M McGuire, Warren C Harvey, William D Bianchi, Alec D Emerling, Erin R Reilly, Vikhyat S Bebarta, Jason J Lopez, Gregory J Zarow, and Jonathan D Auten.
- Department of Emergency Medicine, Naval Medical Center San Diego, San Diego, CA, United States of America.
- Am J Emerg Med. 2022 Apr 1; 54: 58-64.
ObjectivesIntraosseous (IO) access can provide a critical bridge for blood product infusion when peripheral venous access is not obtainable. Successful pressurized IO infusion requires flow rates sufficient to preserve life, but with infusion pressures low enough to avoid clinical complications (e.g., hemolysis, bone damage, fat emboli). However, the optimal method for pressured IO delivery of blood was unknown.MethodsThree trained physicians infused 500 mL of whole blood through a 15-gauge, 45 mm IO catheter into fresh, high bone density cadaveric swine proximal humeri. Participants applied eight different pressure infusion strategies: (1) gravity, (2) pressure bag, (3) pressure bag actively maintained at or above 300 mmHg, (4) hand pump, (5) hand pump with pressure bag, (6) push-pull with 10 mL syringe, (7) push-pull with 60 mL syringe, and a (8) Manual Rapid Infuser in a randomized within-subjects design (30 trials per method, 240 trials total). The primary outcomes of flow rates, mean and peak pressures, and user ratings were contrasted using ANOVA at p < 0.05.ResultsThe Manual Rapid Infuser conferred the highest flow rates (199 ± 3 mL/min) and most favorable user ratings, but also the highest mean and peak pressures. Push-pull conferred the next highest flow rates (67 ± 5 mL/min for 60 mL, 56 ± 2 mL/min for 10 mL) and pressures, with intermediate-to-high user ratings. Hand pump flow rates were essentially identical with (45 ± 4 mL/min) or without (44 ± 3 mL/min) pressure bag, with high user ratings without a pressure bag. Pressure bag and gravity methods conferred low flow rates and user ratings.ConclusionsSome pressured IO infusion methods can achieve flow rates adequate to serve as a resuscitative bridge in the massively hemorrhaged trauma victim, but flow rates and pressures vary greatly across IO pressurized infusion methods. Manual Rapid Infuser and push-pull methods conferred high flow rates but also relatively high pressures, highlighting the importance of using in vivo models in future research to assess the possible clinical complications of using these promising methods. Combined, present findings highlight the importance of studying pressurized IO methods towards preserving the life of the critically injured trauma victim.Published by Elsevier Inc.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.