- Murat Kaya and Nazım Karahan.
- Department of Orthopedics and Traumatology, Marmara University Pendik Training and Research Hospital, Istanbul, Turkey. Electronic address: kayamuratdr@gmail.com.
- Injury. 2022 Jun 1; 53 (6): 1824-1828.
BackgroundIn sawbones with proximal humerus fracture model, three different fixation configurations, Parallel-Straight K-wires, Cross-Straight K-wires and Palm-Tree Method, were biomechanically compared.MethodsA total of 36 anatomical pediatric humerus sawbones models were used. They were divided into three equal groups; parallel fixation with straight K-wires (Group PS), cross fixation with straight K-wires (Group CS), and Palm-Tree Method (Group PT). Models were tested in abduction and torsional at a speed of 0.5 mm/s and a 0-5 mm displacement range. Loading (N) and Stiffness (N/mm) data were calculated and compared statistically.ResultsGroup PS was significantly higher than the other two groups in abduction bending cyclic load values (P<0.001). It was also significantly higher in Group CS than in Group PT (P < 0.001). No significant differences were detected between the three different fixation groups' cyclic torsional load values (p < 0.05).ConclusionThe parallel configuration with straight K-wires will provide a more stable fixation than the cross configuration with straight or Palm-Tree Method in pediatric proximal humeral sawbones fracture modeling.Copyright © 2022. Published by Elsevier Ltd.