Clinical Trial
Performance of the new SmartCardia wireless, wearable oximeter: a comparison with arterial SaO2 in healthy volunteers.
- Francisco Rincon, Julien Pidoux, Srini Murali, and Jean-Jacques Goy.
- Cardiology Department of Clinic Cecil, Av. Ruchonnet 53, 1003, Lausanne, Switzerland.
- BMC Anesthesiol. 2022 Mar 24; 22 (1): 7777.
BackgroundOur goal was to evaluate the performance of a new wearable arm located pulse oximeter.MethodsTwelve volunteers were monitored with three pulse oximeters and underwent desaturation to 70% SaO2. We compared the accuracy of SpO2 reading from the SmartCardia system with SpO2 using two well established devices (Masimo and Nellcor) as reference. Oximetry was performed at different level of oxygen saturation varying from 70 to 100%. Bias, ARMS and precision were evaluated using Bland-Altman plots.ResultsThe mean (SD) differences between SaO2 compared to SpO2 and the devices were as follows: SaO2 versus Masimo 2,12 ± 1,01% (95% CI 1,45 to 2,79), SaO2 versus Nellcor 0,78 ± 0,58% (95% CI - 0,29 to 1,65) and SaO2 versus SmartCardia 0,42 ± 0,24% (95% CI - 0,64 to 1,46). The bias between SmartCardia, Masimo, Nellcor devices and SaO2 was 0.16 (95% CI 0.05 to 0.33) and LoA (level of agreements) 2.96 (95% CI - 2,68 to 2,89) for SmartCardia, 2,02 (95% CI 1,49 to 2,54) and LoA - 6 to 11 for Masimo, and 0,76 (95% CI 0,5 to - 1) and LoA - 3,5 to 5,0 for Nellcor. ARMS for the 70-100% SaO2 range was 1,4 for SmartCardia, 5,0 for Masimo and 2,31 for Nellcor.ConclusionsThe new wireless SmartCardia SpO2 measurement system demonstrated in-line results, bias, ARMS and precision in healthy volunteers, when compared with the gold standard SaO2 and with two well established systems, Masimo and Nellcor.Trial RegistrationThe present trial was prospectively registered at UCSF record (registration number:10-00437), on March 8, 2021.© 2022. The Author(s).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.