• Respiratory care · Jul 2022

    In-Vitro Comparison of Single Limb and Dual Limb Circuit for Aerosol Delivery via Noninvasive Ventilation.

    • Wei Tan, Bing Dai, Dong-Yang Xu, Li-Li Li, and Jie Li.
    • Department of Respiratory and Critical Care Medicine, The First Affiliated Hospital of China Medical University, Shenyang, China.
    • Respir Care. 2022 Jul 1; 67 (7): 807813807-813.

    BackgroundThe effect of single- and dual-limb circuits on aerosol delivery during noninvasive ventilation (NIV) in adult models is unclear.MethodsA noninvasive ventilator equipped with a single-limb circuit or an ICU ventilator equipped with a dual-limb circuit was connected to a simulated lung. Ventilator parameters were adjusted to maintain a tidal volume at ∼500 mL. Aerosol deposition with different placements of a vibrating mesh nebulizer and humidification conditions were compared. Additional experiments by using a non-vented mask or a vented mask were compared in the single-limb circuit only. Aerosol was collected by a disposable filter placed between the simulated lung and the head model (n = 3), and measured by ultraviolet spectrophotometry (276 nm).ResultsThe aerosol deposition varied between 4.12 ± 0.22% and 20.75 ± 0.95%. The greatest aerosol delivery during NIV when using a non-vented mask was found when a vibrating mesh nebulizer was placed between the mask and 15 cm from the exhalation port in the humidified single-limb circuit, and 15 cm from the Y-piece in the inspiratory limb of the humidified dual-limb circuit, and no significant difference of aerosol deposition was found between the two optimal positions (20.03 ± 1.48% vs 18.04 ± 0.93%, respectively; P =.042). There was no difference of aerosol delivery in dry versus humidified circuits, except when a vibrating mesh nebulizer was placed at the humidifier inlet in a dual-limb circuit. When using a vented mask, the aerosol deposition was poor (6.56 ± 0.41 ∼ 8.02 ± 0.39%), regardless of vibrating mesh nebulizer positions and humidification types.ConclusionsDuring NIV, the aerosol delivery was optimal when a vibrating mesh nebulizer was placed between the non-vented mask and 15 cm from the exhalation port in the single-limb circuit or 15 cm from the Y-piece in the inspiratory limb of the dual-limb circuit; no significant difference was found between the two optimal placements. Humidification had little effect on aerosol delivery. Aerosol delivery was poor in the single-limb circuit with a vented mask.Copyright © 2022 by Daedalus Enterprises.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.