Multicenter Study Comparative Study
Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality.
- Christopher M B Fernandes, Andrew Worster, Kevin Eva, Stephen Hill, and Catherine McCallum.
- McMaster University, Hamilton, Ontario, Canada. christopher_fernandes@sympatico.ca
- J Emerg Nurs. 2006 Apr 1;32(2):139-43.
Study ObjectivesIn this study, blood samples from ED patients that were delivered to the laboratory by a pneumatic tube delivery system and by a human courier were compared for timeliness and quality of results.MethodsWe studied all consecutive measurements of serum hemoglobin and potassium ordered from 2 emergency departments of a multisite tertiary care hospital system, one with a pneumatic tube system and the other using human couriers. Turnaround time was measured from the time that the test was ordered by the physician to the time the result was reported on the hospital information system. Hemolysis was measured with use of a standardized, validated method.AnalysisTimes were normalized by log transformation (ln [minutes + 1]), and a comparison of sites was conducted using analysis of variance. Hemolysis rates of the 2 delivery systems were compared by chi2.ResultsThere was no significant difference in hemolysis rate between the 2 methods of delivery (7/121 [5.79%] with a pneumatic tube system and 20/200 [10%] with a human courier). When delivered with a pneumatic tube system, the mean turnaround times (with ranges) for both hemoglobin (33 minutes [4-230]) and potassium (64 [34-208]) were shorter than those delivered by a human courier (43 minutes [3-150] and 72 [28-213], respectively).ConclusionThe use of a pneumatic tube delivery system for transporting blood samples from the emergency department to the laboratory can significantly reduce the turnaround times of results without a reduction in sample quality.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.