• Respiratory care · Sep 2022

    Monitoring of Heated Wire Humidifier Hygrometric Performances With Heater Plate Temperature.

    • François Lellouche, Serge Simard, and Pierre-Alexandre Bouchard.
    • Centre de Recherche de l'Institut Universitaire de Cardiologie et de Pneumologie de Québec, Québec, Québec, Canada. francois.lellouche@criucpq.ulaval.ca.
    • Respir Care. 2022 Sep 1; 67 (9): 114711531147-1153.

    BackgroundUnder-humidification and associated complications may occur with heated humidifiers. Hygrometric performances of heated wire humidifiers are reduced by high ambient and high outlet ventilator temperatures. Currently, there is no reliable monitoring tool to evaluate humidification performances of heated wire humidifiers in the daily practice. We sought to demonstrate the relation between humidity delivered by heated wire humidifiers and different parameters that could be used to monitor humidity of gas delivered to subjects.MethodsOn a bench test, we measured heater plate temperature, inlet chamber temperature, and delivered humidity with MR850 system. Temperature displayed on the humidifier was also recorded. The measurements were performed at different ambient temperatures and five minute ventilation levels (5, 7.5, 10, 12.5, 15 L/min). Inlet chamber temperatures varied from 20-40°C. In each condition, hygrometric measurements with the psychrometric method were performed at steady state.ResultsWe performed 279 measurements at steady state including all conditions. We found a good correlation between heater plate temperature and absolute humidity delivered (R2 = 0.82). This relationship was hardly affected by ambient temperature, but minute ventilation had more effect. For different minute ventilations, the correlation between heater plate temperature and absolute humidity delivered was very good with coefficient of determination R2 from 0.87-0.98. Heater plate temperature > 62°C was a good predictor of absolute humidity delivered > 30 mg H2O/L (area under the curve = 0.96, sensibility 79%, specificity 94%). No correlation existed between humidity delivered and the outlet chamber temperature (displayed on the humidifier).ConclusionsIn this bench study, we have shown a good correlation between heater plate temperature and humidity delivered with a heated wire humidifier. This means that a "hidden" hygrometer is built into the heated wire humidifier. Heater plate temperature should be used as a surrogate of humidity to improve the humidification monitoring.Copyright © 2022 by Daedalus Enterprises.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.