- Lien-Chen Wu, Yueh-Ying Hsieh, Fon-Yih Tsuang, Yueh-Feng Chiang, and Chang-Jung Chiang.
- Department of Orthopedics, Shuang Ho Hospital, Taipei Medical University, New Taipei City 23561, Taiwan.
- Medicina (Kaunas). 2022 Jun 15; 58 (6).
AbstractBackground and Objectives: Using an appropriate torque to tighten set screws ensures the long-term stability of spinal posterior fixation devices. However, the recommended torque often varies between different devices and some devices do not state a recommended torque level. The purpose of this study is to evaluate the effect of set screw torque on the overall construct stability and fatigue life. Materials and Methods: Two commercial pedicle screw systems with different designs for the contact interface between the set screw and rod (Group A: plane contact, Group B: line contact) were assembled using torque wrenches provided with the devices to insert the set screws and tighten to the device specifications. The axial gipping capacity and dynamic mechanical stability of each bilateral construct were assessed in accordance with ASTM F1798 and ASTM F1717. Results: Increasing or decreasing the torque on the set screw by 1 Nm from the recommended level did not have a significant effect on the axial gripping capacity or fatigue strength of Group A (p > 0.05). For Group B, over-tightening the set screw by 1 Nm did cause a significant reduction in the fatigue strength. Conclusions: Excessive torque can damage the rod surface and cause premature failure. When insertion using a manual driver is preferred, a plane contact interface between the set screw and rod can reduce damage to the rod surface when the set screw is over-torqued.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.