    BackgroundThe neural integrity monitor (NIM) electromyogram (EMG) endotracheal (ET) tube is a widely used device to monitor neural response through muscle activity. It is helpful in surgical procedures with high risk of damaging delicate structures in the head and neck. This case provides a thorough analysis of an adverse event that was encountered in the operating room, which others can hopefully learn from.Case PresentationWe are reporting a case in which a patient undergoing hemithyroidectomy had experienced an adverse event using the Medtronic NIM™ EMG endotracheal tube. After successful induction and intubation, confirming the proper positioning of the electrode wires was necessary before the incision could be made. Upon reexamination, the patient suddenly became difficult to ventilate with increased peak airway pressure, decreased tidal volume, and end tidal CO2. This episode lasted approximately 15 min and the patient's condition remained stable despite low tidal volumes. The problem was unexpectedly resolved upon deflation of the cuff of the ET tube.ConclusionsThere are several similar reports of these endotracheal tubes causing obstruction, especially those in which overinflation of the cuff caused cuff herniation and blockage of the Murphy eye and the bevel. It is currently believed that the design of this tube allowed for the obstruction to occur. The patient's short body habitus may have also been a small contributing factor. The distance that the electrodes must sit within the vocal cords to the tip of the bevel is longer in this type of ET tube compared with a standard ET tube. The distance from the true vocal cords to where the cuff sits in the trachea is also greater in this model NIM EMG tube. There was no confirmation of the exact obstructive process that took place, however, confirming the tube and cuff positioning would have been optimal.© 2022. The Author(s).

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…