Paediatric anaesthesia · Oct 2022
Observational StudyComparison of Two Point-of-Care Measurements of Hemoglobin against Standard Laboratory Values during Pediatric Liver Transplantation: A Retrospective Review.
- Bethany L Morris, John Hajduk, Eric C Cheon, Heather A Ballard, and Nicholas E Burjek.
- Division of Pediatric Anesthesiology, Children's Hospital Colorado, Aurora, Colorado, USA.
- Paediatr Anaesth. 2022 Oct 1; 32 (10): 1121-1128.
BackgroundPoint-of-care hemoglobin testing devices play an important role in intraoperative anesthetic management where significant hemorrhage is anticipated; however, the reliability of these devices has not been examined in the context of pediatric liver transplantation. In this retrospective observational study, we aimed to determine whether 95% of results from two point-of-care hemoglobinometers, the HemoCue and iSTAT, would fall within a difference of ±1 g/dl, our a priori-defined clinically acceptable level of agreement, of the hemoglobin measures on a core laboratory complete blood count.MethodsWe retrospectively collected data from 70 patients presenting for a liver transplant at a single center, tertiary care pediatric hospital over a 3.5-year period. We analyzed 92 contemporaneous pairs of hemoglobin values from the HemoCue and complete blood count, and 252 pairs of hemoglobin values from the iSTAT and complete blood count. Agreement between the point-of-care devices and complete blood count was assessed using Bland-Altman analysis, which was the primary outcome. Secondary analyses included an error grid analysis and Cohen's kappa statistic.ResultsBoth point-of-care devices underestimated complete blood count hemoglobin values and neither device satisfied our a priori-defined clinically acceptable level of agreement that 95% of values would fall within ±1 g/dl of the complete blood count measurement. The mean difference [limits of agreement] of the HemoCue was 0.4 g/dl (p < .001) [-0.9 to 1.6 g/dl] and of the iSTAT was 0.6 g/dl (p < .001) [-1.4 to 2.6 g/dl]. Secondary error grid analysis revealed that neither device performed well enough to replace a complete blood count at critical thresholds of hemoglobin values.ConclusionsWhile the HemoCue and iSTAT contribute information in a timely manner during dynamic intraoperative situations, there is significant imprecision compared to the gold standard complete blood count. If clinical stability allows, the results of these point-of-care hemoglobinometers should be confirmed with a complete blood count, rather than being used as the sole factor in determining transfusion needs during pediatric liver transplantation.© 2022 John Wiley & Sons Ltd.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.