- Frederik S Stecher, Jan-Aage Olsen, Ronald E Stickney, and Lars Wik.
- National Competence Centre of Emergency Medicine, Institute for Experimental Medical Research, Ulleval University Hospital, 0027 Oslo, Norway.
- Resuscitation. 2008 Dec 1;79(3):432-7.
IntroductionThere is a need to measure cardiopulmonary resuscitation (CPR) in order to document whether ambulance personnel follow CPR guidelines. Our goal was to do this using defibrillator technology based on changes in transthoracic impedance (TTI) produced by chest compressions and ventilations.Methods122 incidents of out-of-hospital cardiac arrest between May 2003 and February 2004 were analysed based on data recorded from defibrillators in Oslo EMS. New software was used to analyze chest compressions and ventilations based on changes in thoracic impedance between the defibrillator pads, as well as ECG and other event data.ResultsIn total, 25+/-14% (varying from 76% to 3%) of the time chest compressions were not performed on patients without spontaneous circulation (NFR=No Flow Ratio). When adjusting for time spent on analysis of ECG, pulse check and defibrillation, NFR was 20+/-13% (varying from 70% to 3%). Mean compressions delivered per minute was 87+/-16 and the compression rate during active compressions was 117+/-9min(-1). Individual variation was 31-117min(-1) (mean) and 95-144min(-1) (active periods). A mean of 14+/-3ventilations/min was recorded, varying from 8 to 26min(-1). Compared with the rest of the episode, the first 5min had a significantly higher proportion of time without chest compressions; 30+/-17% (p<0.001) and significantly lower mean compression and ventilation rates; 80+/-19min(-1) and 12+/-4min(-1), respectively (p<0.001 in both cases).ConclusionsCore CPR values can be measured from TTI signals by using a standard defibrillator and new software. NFR was 25% (20% adjusted) with great rescuer variability.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.