J Clin Monit Comput · Feb 2023
Is it feasible to measure intra-abdominal pressure using a balloon-tipped rectal catheter? Results of a validation study.
- Anneleen S Staelens, Ann Heymans, Sigrid Christiaens, Niels Van Regenmortel, Wilfried Gyselaers, and MalbrainManu L N GMLNGFirst Department of Anaesthesiology and Intensive Therapy, Medical University of Lublin, Jaczewskiego Street 8, 20-954, Lublin, Poland. manu.malbrain@telenet.be.Medical Data Management, Medaman, 2440, Geel, Belgium. manu.malbrain@tel.
- Department Obstetrics and Gynaecology, Ziekenhuis Oost Limburg, Genk, Belgium.
- J Clin Monit Comput. 2023 Feb 1; 37 (1): 287296287-296.
AbstractThe gold standard to measure intra-abdominal pressure (IAP) is intra-vesical measurement via the urinary bladder. However, this technique is restricted in ambulatory settings because of the risk of iatrogenic urinary tract infections. Rectal IAP measurements (IAPrect) may overcome these limitations, but requires validation. This validation study compares the IAPrect technique against gold standard intra-vesical IAP measurements (IAPves). IAPrect using an air-filled balloon catheter and IAPves using Foley Manometer Low Volume were measured simultaneously in sedated and ventilated patients. Measurements were performed twice in different positions (supine and HOB 45° elevated head of bed) and with an external abdominal pressure belt. Sixteen patients were included. Seven were not eligible for analysis due to unreliable IAPrect values. IAPrect was significantly higher than IAPves for all body positions (p < 0.01) and the correlation between IAPves and IAPrect was poor and not significant in each position (p ≥ 0.25, R2 < 0.6, Lin's CCC < 0.8, bias - 8.1 mmHg and precision of 5.6 mmHg with large limits of agreement between - 19 to 2.9 mmHg, high percentage error 67.3%, and low concordance 86.2%). Repeatability of IAPrect was not reliable (R = 0.539, p = 0.315). For both techniques, measurements with the external abdominal pressure belt were significantly higher compared to those without (p < 0.03). IAPrect has important shortcomings making IAP estimation using a rectal catheter unfeasible because the numbers cannot be trusted nor validated.© 2022. The Author(s).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.