-
- Tobias Teucke, F Maurer, L M Müller-Wirtz, T Volk, D I Sessler, and S Kreuer.
- CBR - Center of Breath Research, Department of Anaesthesiology, Intensive Care and Pain Therapy, Faculty of Medicine, Saarland University Medical Center, Saarland University, 66421, Homburg, Saar, Germany. t.t.teucke@gmail.com.
- J Clin Monit Comput. 2023 Apr 1; 37 (2): 493500493-500.
AbstractThe bedside Exhaled Drug MONitor - EDMON measures exhaled propofol in ppbv every minute based on multi-capillary column - ion mobility spectrometry (MCC-IMS). The MCC pre-separates gas samples, thereby reducing the influence of the high humidity in human breath. However, preliminary analyses identified substantial measurement deviations between dry and humid calibration standards. We therefore performed an analytical validation of the EDMON to evaluate the influence of humidity on measurement performance. A calibration gas generator was used to generate gaseous propofol standards measured by an EDMON device to assess linearity, precision, carry-over, resolution, and the influence of different levels of humidity at 100% and 1.7% (without additional) relative humidity (reference temperature: 37°C). EDMON measurements were roughly half the actual concentration without additional humidity and roughly halved again at 100% relative humidity. Standard concentrations and EDMON values correlated linearly at 100% relative humidity (R²=0.97). The measured values were stable over 100min with a variance ≤ 10% in over 96% of the measurements. Carry-over effects were low with 5% at 100% relative humidity after 5min of equilibration. EDMON measurement resolution at 100% relative humidity was 0.4 and 0.6 ppbv for standard concentrations of 3 ppbv and 41 ppbv. The influence of humidity on measurement performance was best described by a second-order polynomial function (R²≥0.99) with influence reaching a maximum at about 70% relative humidity. We conclude that EDMON measurements are strongly influenced by humidity and should therefore be corrected for sample humidity to obtain accurate estimates of exhaled propofol concentrations.© 2022. The Author(s).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.