- Amanda J Nickel, Howard B Panitch, Joseph M McDonough, Etze Chotzoglou, and Julian L Allen.
- Department of Respiratory Care, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania. nickelaj@chop.edu.
- Respir Care. 2022 Nov 1; 67 (11): 140514121405-1412.
BackgroundIntrinsic PEEP during mechanical ventilation occurs when there is insufficient time for expiration to functional residual capacity before the next inspiration, resulting in air trapping. Increased expiratory resistance (RE), too rapid of a patient or ventilator breathing rate, or a longer inspiratory to expiratory time ratio (TI/TE) can all be causes of intrinsic PEEP. Intrinsic PEEP can result in increased work of breathing and patient-ventilator asynchrony (PVA) during patient-triggered breaths. We hypothesized that the difference between intrinsic PEEP and ventilator PEEP acts as an inspiratory load resulting in trigger asynchrony that needs to be overcome by increased respiratory muscle pressure (Pmus).MethodsUsing a Servo lung model (ASL 5000) and LTV 1200 ventilator in pressure control mode, we developed a passive model demonstrating how elevated RE increases intrinsic PEEP above ventilator PEEP. We also developed an active model investigating the effects of RE and intrinsic PEEP on trigger asynchrony (expressed as percentage of patient-initiated breaths that failed to trigger). We then studied if trigger asynchrony could be reduced by increased Pmus.ResultsIntrinsic PEEP increased significantly with increasing RE (r = 0.97, P = .006). Multivariate logistic regression analysis showed that both RE and negative Pmus levels affect trigger asynchrony (P < .001).ConclusionsA passive lung model describes the development of increasing intrinsic PEEP with increasing RE at a given ventilator breathing rate. An active lung model shows how this can lead to trigger asynchrony since the Pmus needed to trigger a breath is greater with increased RE, as the inspiratory muscles must overcome intrinsic PEEP. This model will lend itself to the study of intrinsic PEEP engendered by a higher ventilator breathing rate, as well as higher TI/TE, and will be useful in ventilator simulation scenarios of PVA. The model also suggests that increasing ventilator PEEP to match intrinsic PEEP can improve trigger asynchrony through a reduction in RE.Copyright © 2022 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.