J Clin Monit Comput · Jun 2023
Controlled Clinical TrialAgreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study.
- Ilonka N De Keijzer, Dario Massari, Caren K Niezen, BokkersReinoud P HRPHDepartment of Radiology, Medical Imaging Center, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands., Jaap Jan Vos, and ScheerenThomas W LTWLDepartment of Anesthesiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.Currently: Senior Director Medical Affairs, Edwards Lifesciences, Irvine, CA, USA..
- Department of Anesthesiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands. i.n.de.keijzer@umcg.nl.
- J Clin Monit Comput. 2023 Jun 1; 37 (3): 805814805-814.
PurposeO3® Regional Oximetry (Masimo Corporation, California, USA) is validated for cerebral oximetry. We aimed to assess agreement of somatic and renal near-infrared spectroscopy with reference blood samples.MethodsO3 sensors were placed bilaterally on the quadriceps and flank of 26 healthy volunteers. A stepped, controlled hypoxia sequence was performed by adding a mixture of nitrogen and room air to the breathing circuit. O3-derived oxygen saturation values were obtained at baseline and at six decremental saturation levels (5% steps). Blood samples (radial artery, iliac vein (somatic reference) and renal vein) were obtained at each step. Reference values were calculated as: 0.7 × venous saturation + 0.3 × arterial saturation. The agreement between O3-derived values with blood reference values was assessed by calculating root-mean-square error accuracy and Bland-Altman plots.ResultsThe root-mean-square error accuracy was 6.0% between quadriceps oxygen saturation and somatic reference values. The mean bias was 0.8%, with limits of agreement from -7.7 to 9.3%. These were 5.1% and 0.6% (-8.3 to 9.5%) for flank oxygen saturation and somatic reference values, respectively, and 7.7% and -4.9% (-15.0 to 5.2%) for flank oxygen saturation and renal reference values. The kidney depth was 3.1 ± 0.9 cm below the skin.ConclusionO3 regional oximetry can be used on the quadriceps and flank to monitor somatic saturation, yet has a saturation-level dependent bias. O3-derived values obtained at the flank underestimated renal reference values. Additionally, it is unlikely that the flank sensors did directly measure renal tissue.Trial RegistrationClinicaltrials.gov (NCT04584788): registered October 6th, 2020.© 2022. The Author(s).
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.