• Eur J Anaesthesiol · Mar 2008

    Accuracy of cardiac output measurements with pulse contour analysis (PulseCO) and Doppler echocardiography during off-pump coronary artery bypass grafting.

    • C Missant, S Rex, and P F Wouters.
    • Katholieke Universiteit Leuven, University Hospitals, Department of Acute Medical Sciences-Anaesthesiology, U. Z. Gasthuisberg, Leuven, Belgium.
    • Eur J Anaesthesiol. 2008 Mar 1;25(3):243-8.

    Background And ObjectiveDuring off-pump coronary bypass grafting, surgical manipulation and dislocation of the heart may cause cardiovascular instability. Monitoring of cardiac output facilitates intraoperative haemodynamic management but pulmonary artery catheters are often considered too invasive. Pulse contour analysis and transoesophageal echocardiography could serve as alternatives, but there is controversy about their accuracies. We validated pulse contour analysis using a standard radial arterial catheter (PulseCO) and aortic Doppler flowmetry with transoesophageal echocardiography in patients undergoing off-pump coronary bypass surgery. Pulmonary arterial thermodilution served as the reference technique.MethodsIn 20 patients undergoing off-pump coronary bypass, cardiac output was measured with bolus thermodilution (COTD), pulse contour analysis (COPC), and transoesophageal echocardiography (COecho) at fixed time intervals during the procedure. Data were compared using linear regression and Bland-Altman analysis. At the end of the procedure, dobutamine was infused at a rate of 2.5 microg kg(-1) min(-1) in six patients to study the agreement between methods in quantifying changes in cardiac output.ResultsComparison between COPC and COTD showed a bias +/- limits of agreement of -0.03 +/- 1.30 L min(-1) (mean error 29%). Doppler echocardiography was not always feasible when the heart was displaced from the oesophagus and had lower accuracy: bias +/- limits of agreement vs. COTD was 0.45 +/- 1.93 (mean error 43%). Increases in cardiac output induced by dobutamine were well quantified both by pulse contour analysis (COPC = 0.76 x COTD + 0.58; r(2) = 0.65) and Doppler, although the latter tended to overestimate these changes (COecho = 1.58 x COTD - 0.13; r(2) = 0.53).ConclusionsCalibrated pulse contour analysis using the PulseCO system is an acceptable technique to measure cardiac output non-invasively in off-pump coronary bypass patients. Doppler echocardiography performs less well and is not always feasible with transoesophageal echocardiography when the heart is displaced.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.