Comparative StudyAmbient hemolysis and activation of coagulation is different between HeartMate II and HeartWare left ventricular assist devices.
- Ingvild Birschmann, Marcus Dittrich, Thomas Eller, Bettina Wiegmann, Armin J Reininger, Ulrich Budde, and Martin Strüber.
- Department of Hematology, Hemostasis, Oncology and Stem Cell Transplantation, Hannover Medical School, Hannover, Germany. Electronic address: ibirschmann@hdz-nrw.de.
- J. Heart Lung Transplant. 2014 Jan 1;33(1):80-7.
BackgroundThromboembolic and bleeding events in patients with a left ventricular assist device (LVAD) are still a major cause of complications. Therefore, the balance between anti-coagulant and pro-coagulant factors needs to be tightly controlled. The principle hypothesis of this study is that different pump designs may have an effect on hemolysis and activation of the coagulation system. Referring to this, the HeartMate II (HMII; Thoratec Corp, Pleasanton, CA) and the HeartWare HVAD (HeartWare International Inc, Framingham, MA) were investigated.MethodsFor 20 patients with LVAD support (n = 10 each), plasma coagulation, full blood count, and clinical chemistry parameters were measured. Platelet function was monitored using platelet aggregometry, platelet function analyzer-100 system ( Siemens, Marburg, Germany), vasodilator-stimulated phosphoprotein phosphorylation assay, immature platelet fraction, platelet-derived microparticles, and von Willebrand diagnostic.ResultsAcquired von Willebrand syndrome could be detected in all patients. Signs of hemolysis, as measured by lactate dehydrogenase levels (mean, 470 U/liter HMII, 250 U/liter HVAD; p < 0.001), were more pronounced in the HMII patients. In contrast, D-dimer analysis indicated a significantly higher activation of the coagulation system in HVAD patients (mean, 0.94 mg/liter HMII, 2.01 mg/liter HVAD; p < 0.01). The efficacy of anti-platelet therapy using clopidogrel was not sufficient in more than 50% of the patients.ConclusionsOur results support the finding that all patients with rotary blood pumps suffered from von Willebrand syndrome. In addition, a distinct footprint of effects on hemolysis and the coagulation system can be attributed to different devices. As a consequence, the individual status of the coagulation system needs to be controlled in long-term patients.© 2013 Published by International Society for the Heart and Lung Transplantation on behalf of International Society for Heart and Lung Transplantation.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.