-
- Marie-Anne Blanchet, Gabriel Mercier, Antoine Delobel, Emi Nayet, Pierre-Alexandre Bouchard, Serge Simard, Erwan L'Her, Richard D Branson, and François Lellouche.
- Department of Medicine, Centre de recherche, Institut Universitaire de Cardiologie et de Pneumologie de Québec, Université Laval, Québec, Canada.
- Respir Care. 2023 May 1; 68 (5): 565574565-574.
BackgroundAn accurate SpO2 value is critical in order to optimally titrate oxygen delivery to patients and to follow oxygenation guidelines. Limited prospective data exist on real-world performance of pulse oximeters in critically ill patients. The objective of this study was to assess accuracy and bias of the SpO2 values measured by several oximeters in hospitalized subjects.MethodsWe included stable adults in the ICU with an arterial catheter in place. Main exclusion criteria were poor SpO2 signal and SpO2 > 96%. In each subject, we simultaneously evaluated 4 oximeters: Nonin (Plymouth, Minnesota) embedded in the FreeO2 device (OxyNov, Québec City, Québec, Canada), Masimo (Radical-7, Masimo, Irvine, California), Philips (FAST, Philips, Amsterdam, the Netherlands), and Nellcor (N-600, Medtronic, Minneapolis, Minnesota). Arterial blood gases were drawn and simultaneously each oximeters' SpO2 values were collected. SpO2 values were compared to the reference (arterial oxygen saturation [SaO2 ] value) to determine bias and accuracy. The ability for oximeters to detect hypoxemia and the impact of oximeters on oxygen titration were evaluated.ResultsWe included 193 subjects (153 male, mean age 66 y) in whom 211 sets of measurements were performed. The skin pigmentation evaluated by Fitzpatrick scale showed 96.2% of subjects were light skin (types 1 and 2). One oximeter overestimated SaO2 (Philips, +0.9%), whereas the 3 others underestimated SaO2 (Nonin -3.1%, Nellcor -0.3%, Masimo -0.2%). SaO2 was underestimated with Nonin oximeter in 91.3% of the cases, whereas it was overestimated in 55.2% of the cases with Philips oximeter. Moderate hypoxemia (SaO2 86-90% or PaO2 55-60 mm Hg) was detected in 92, 33, 42, and 11% of the cases with Nonin, Nellcor, Masimo, and Philips, respectively.ConclusionsWe found significant bias and moderate accuracy between the tested oximeters and the arterial blood gases in the studied population. These discrepancies may have important clinical impact on the detection of hypoxemia and management of oxygen therapy.Copyright © 2023 by Daedalus Enterprises.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.