• Neurosurgery · Feb 2023

    First Experience With Postoperative Transcranial Ultrasound Through Sonolucent Burr Hole Covers in Adult Hydrocephalus Patients.

    • Ryan P Lee, Michael Meggyesy, Jheesoo Ahn, Christina Ritter, Ian Suk, A Judit Machnitz, Judy Huang, Chad Gordon, Henry Brem, and Mark Luciano.
    • Department of Neurosurgery, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
    • Neurosurgery. 2023 Feb 1; 92 (2): 382390382-390.

    BackgroundManaging patients with hydrocephalus and cerebrospinal fluid (CSF) disorders requires repeated head imaging. In adults, it is typically computed tomography (CT) or less commonly magnetic resonance imaging (MRI). However, CT poses cumulative radiation risks and MRI is costly. Ultrasound is a radiation-free, relatively inexpensive, and optionally point-of-care alternative, but is prohibited by very limited windows through an intact skull.ObjectiveTo describe our initial experience with transcutaneous transcranial ultrasound through sonolucent burr hole covers in postoperative hydrocephalus and CSF disorder patients.MethodsUsing cohort study design, infection and revision rates were compared between patients who underwent sonolucent burr hole cover placement during new ventriculoperitoneal shunt placement and endoscopic third ventriculostomy over the 1-year study time period and controls from the period 1 year before. Postoperatively, trans-burr hole ultrasound was performed in the clinic, at bedside inpatient, and in the radiology suite to assess ventricular anatomy.ResultsThirty-seven patients with sonolucent burr hole cover were compared with 57 historical control patients. There was no statistically significant difference in infection rates between the sonolucent burr hole cover group (1/37, 2.7%) and the control group (0/57, P = .394). Revision rates were 13.5% vs 15.8% (P = 1.000), but no revisions were related to the burr hole or cranial hardware.ConclusionTrans-burr hole ultrasound is feasible for gross evaluation of ventricular caliber postoperatively in patients with sonolucent burr hole covers. There was no increase in infection rate or revision rate. This imaging technique may serve as an alternative to CT and MRI in the management of select patients with hydrocephalus and CSF disorders.Copyright © Congress of Neurological Surgeons 2022. All rights reserved.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.