Critical care medicine · Feb 2023
Observational StudyProspective Real-Time Validation of a Lung Ultrasound Deep Learning Model in the ICU.
- Chintan Dave, Derek Wu, Jared Tschirhart, Delaney Smith, Blake VanBerlo, Jason Deglint, Faraz Ali, Rushil Chaudhary, Bennett VanBerlo, Alex Ford, Marwan A Rahman, Joseph McCauley, Benjamin Wu, Jordan Ho, Brian Li, and Robert Arntfield.
- Division of Critical Care Medicine, Western University, London, ON, Canada.
- Crit. Care Med. 2023 Feb 1; 51 (2): 301309301-309.
ObjectivesTo evaluate the accuracy of a bedside, real-time deployment of a deep learning (DL) model capable of distinguishing between normal (A line pattern) and abnormal (B line pattern) lung parenchyma on lung ultrasound (LUS) in critically ill patients.DesignProspective, observational study evaluating the performance of a previously trained LUS DL model. Enrolled patients received a LUS examination with simultaneous DL model predictions using a portable device. Clip-level model predictions were analyzed and compared with blinded expert review for A versus B line pattern. Four prediction thresholding approaches were applied to maximize model sensitivity and specificity at bedside.SettingAcademic ICU.PatientsOne-hundred critically ill patients admitted to ICU, receiving oxygen therapy, and eligible for respiratory imaging were included. Patients who were unstable or could not undergo an LUS examination were excluded.InterventionsNone.Measurements And Main ResultsA total of 100 unique ICU patients (400 clips) were enrolled from two tertiary-care sites. Fifty-six patients were mechanically ventilated. When compared with gold standard expert annotation, the real-time inference yielded an accuracy of 95%, sensitivity of 93%, and specificity of 96% for identification of the B line pattern. Varying prediction thresholds showed that real-time modification of sensitivity and specificity according to clinical priorities is possible.ConclusionsA previously validated DL classification model performs equally well in real-time at the bedside when platformed on a portable device. As the first study to test the feasibility and performance of a DL classification model for LUS in a dedicated ICU environment, our results justify further inquiry into the impact of employing real-time automation of medical imaging into the care of the critically ill.Copyright © 2023 by the Society of Critical Care Medicine and Wolters Kluwer Health, Inc. All Rights Reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.