• Reg Anesth Pain Med · Oct 2023

    Accuracy of injection pressure measurement at peripheral nerves using high-resolution 40 MHz ultrasound in an anesthetized porcine model.

    • Graeme A McLeod, Alistair Cowie, Amy Sadler, Fiona Watson, Paul Wasik, and Miguel Angel Reina.
    • Department of Anaesthesia, Ninewells Hospital, Dundee, UK g.a.mcleod@dundee.ac.uk.
    • Reg Anesth Pain Med. 2023 Oct 1; 48 (10): 501507501-507.

    BackgroundFluid injection pressure measurement is promoted as a marker of needle tip position that discriminates between tissue layers. However, clinical ultrasound has insufficient resolution to identify the exact position of the needle tip. Our primary objective was to use 40 MHz ultrasound in anesthetized pigs in order to precisely locate the tip of the needle and measure opening injection pressure in muscle, at epineurium and in subepineurium.MethodsWe surgically exposed the axillae of four anesthetized pigs. Two operators placed a 40 MHz ultrasound transducer over the pectoral muscle and imaged axillary, median and radial nerves. Injections (0.5 mL) were randomized to in-plane and out-of-plane needle trajectories and flow rates of 1, 6 and 12 mL/min.ResultsWe identified 541 fascicles in 23 nerves. The ratio of fascicle area to nerve area remained constant at ~0.30 for all nerves. Axillary nerves were smaller than median and radial nerves, difference in diameter (95% CI) 1.61 (0.87 to 2.36) mm, p<0.001 and 1.59 (0.82 to 2.36) mm, p=0.001, respectively. Axillary nerves had less fascicles per nerve than median nerves, difference 7.63 (2.43 to 12.83) and radial nerves, difference 9.02 (3.64 to 14.40). We visualized the circumneurium and injection within the subcircumneural compartment. Intraneural injection increased nerve area (SD) from 5.7 (2.2) mm2 to 13.7 (5.5) mm2, difference 8.0 (5.4-10.6) mm2, p<0.001. Mean injection pressure was greater in subepineurium compared with muscle, geometric ratio 2.29 (1.30 to 4.10), p<0.001; and greater on epineurium compared with muscle, geometric ratio 1.73 (1.03 to 3.00), p=0.01. Twenty-two out of 23 injections in muscle, 14 out of 23 injections at epineurium and 11 out of 22 injections in subepineurium were <138 kPa (20 psi).ConclusionNeedle tip position was not discernible using pressure monitoring. The circumneurium and subcircumneural injection compartment were observed but not intrafascicular injection.© American Society of Regional Anesthesia & Pain Medicine 2023. No commercial re-use. See rights and permissions. Published by BMJ.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.