Randomized Controlled Trial
In Vitro Testing and Clinical Handling of a Novel Implant Positioning Technology for Proximal Humeral Plating.
- Markus Windolf, Dominik Knierzinger, Stefaan Nijs, An Sermon, Michael Blauth, RichardsRobert GeoffRG0000-0002-7778-2480AO Research Institute Davos, 7270 Davos, Switzerland., and Jan Buschbaum.
- AO Research Institute Davos, 7270 Davos, Switzerland.
- Medicina (Kaunas). 2023 Feb 23; 59 (3).
Background And ObjectivesFractures of the proximal humerus are common, particularly in elderly populations. Anatomical locking plates target stabilization with a multitude of screws spanning into the humeral head. Sound implant placement and screw length determination are key for a successful clinical outcome but are difficult to obtain from planar X-rays. A novel implant positioning technology for proximal humerus plating (Xin1) outputs screw lengths suggestions and plate position based on hole projections in conventional X-ray images. This study investigated the performance of a prototype Xin1 system in a postmortem (in vitro) experiment as well as in a clinical handling test.Materials And MethodsFor in vitro testing, twelve shoulders from six anatomical specimens were randomized into two groups to compare the Xin1 technique to the conventional operation in terms of surgical precision, procedure time and X-ray exposure. For the clinical trial, 11 patients undergoing plating of the proximal humerus were included. The aim was to investigate clinical handling of the Xin1 marker clip and to retrospectively evaluate the system performance in a real-life fracture situation. Image pairs before and after insertion of the proximal screws were retrospectively processed to investigate the influence of potential bone fragment shifts on the system output.ResultsIn the postmortem experiment, the use of the system significantly improved the surgical precision (52% error reduction), procedure time (38% shorter) and radiation exposure (64% less X-rays). Clinical handling demonstrated seamless embedding of the marker clip into existing clinical workflows without adverse events reported. Retrospective X-ray analysis on six eligible patients revealed differences in the calculated screw lengths of ≤2 mm before and after screw insertion for five patients. In one patient, the screw lengths differed up to 8 mm, which might indicate displacement of the head fragment.ConclusionsResults suggest a strong potential of the Xin1 assistance technology to enhance the surgical procedure and patient outcomes in the rising incidence of osteoporotic humeral fractures. Robust performance in a real-life fracture situation was observed. In-depth validation of the system is, however, needed before placing it into clinical practice.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:

- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.