• Crit Care · Mar 2023

    Randomized Controlled Trial

    Impact on the ability of healthcare professionals to correctly identify patient-ventilator asynchronies of the simultaneous visualization of estimated muscle pressure curves on the ventilator display: a randomized study (Pmus study).

    • Daniel Oliveira Silva, Patrícia Nery de Souza, de Araujo SousaMayson LaercioMLInterdepartmental Division of Critical Care Medicine, St. Michael's Hospital, Toronto, Canada., Caio Cesar Araujo Morais, Juliana Carvalho Ferreira, Marcelo Alcantara Holanda, Wellington Pereira Yamaguti, JuniorLaerte PastoreLPIntensive Care Unit, Hospital Sírio-Libanes, São Paulo, Brazil., and CostaEduardo Leite VieiraELVDisciplina de Pneumologia, Heart Institute (Incor), Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, São Paulo, Brazil.Research and Education Institute, Hospital Sírio-Libanes, São Paulo, Brazil..
    • Intensive Care Unit, Hospital Sírio-Libanes, São Paulo, Brazil.
    • Crit Care. 2023 Mar 30; 27 (1): 128128.

    BackgroundPatient-ventilator asynchronies are usually detected by visual inspection of ventilator waveforms but with low sensitivity, even when performed by experts in the field. Recently, estimation of the inspiratory muscle pressure (Pmus) waveforms through artificial intelligence algorithm has been proposed (Magnamed®, São Paulo, Brazil). We hypothesized that the display of these waveforms could help healthcare providers identify patient-ventilator asynchronies.MethodsA prospective single-center randomized study with parallel assignment was conducted to assess whether the display of the estimated Pmus waveform would improve the correct identification of asynchronies in simulated clinical scenarios. The primary outcome was the mean asynchrony detection rate (sensitivity). Physicians and respiratory therapists who work in intensive care units were randomized to control or intervention group. In both groups, participants analyzed pressure and flow waveforms of 49 different scenarios elaborated using the ASL-5000 lung simulator. In the intervention group the estimated Pmus waveform was displayed in addition to pressure and flow waveforms.ResultsA total of 98 participants were included, 49 per group. The sensitivity per participant in identifying asynchronies was significantly higher in the Pmus group (65.8 ± 16.2 vs. 52.94 ± 8.42, p < 0.001). This effect remained when stratifying asynchronies by type.ConclusionsWe showed that the display of the Pmus waveform improved the ability of healthcare professionals to recognize patient-ventilator asynchronies by visual inspection of ventilator tracings. These findings require clinical validation.Trial RegistrationClinicalTrials.gov: NTC05144607. Retrospectively registered 3 December 2021.© 2023. The Author(s).

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.