• Anesthesiology · Aug 2023

    Comparative Study

    Comparison of a Modern Digital Mechanomyograph to a Mechanomyograph Utilizing an Archival Grass Force Transducer.

    • Kelly E Michaelsen, Srdjan Jelacic, Sharon T Nguyen, Kishanee J Haththotuwegama, Kei Togashi, and Andrew Bowdle.
    • Department of Anesthesiology and Pain Medicine, University of Washington, Seattle, Washington.
    • Anesthesiology. 2023 Aug 1; 139 (2): 164172164-172.

    BackgroundMechanomyography is the traditional gold standard research technique for quantitative assessment of neuromuscular blockade. Mechanomyography directly measures the isometric force generated by the thumb in response to ulnar nerve stimulation. Researchers must construct their own mechanomyographs since commercial instruments are no longer available. A mechanomyograph was constructed, and its performance was compared against an archival mechanomyography system from the 1970s that utilized an FT-10 Grass force transducer, hypothesizing that train-of-four ratios recorded on each device would be equivalent.MethodsA mechanomyograph was constructed using 3D-printed components and modern electronics. An archival mechanomyography system was assembled from original components, including an FT-10 Grass force transducer. Signal digitization for computerized data collection was utilized instead of the original paper strip chart recorder. Both devices were calibrated with standard weights to demonstrate linear voltage response curves. The mechanomyographs were affixed to opposite arms of patients undergoing surgery, and the train-of-four ratio was measured during the onset and recovery from rocuronium neuromuscular blockade.ResultsCalibration measurements exhibited a positive linear association between voltage output and calibration weights with a linear correlation coefficient of 1.00 for both mechanomyography devices. The new mechanomyograph had better precision and measurement sensitivity than the archival system: 5.3 mV versus 15.5 mV and 1.6 mV versus 5.7 mV, respectively (P < 0.001 for both). A total of 767 pairs of train-of-four ratio measurements obtained from eight patients had positive linear association (R 2 = 0.94; P < 0.001). Bland-Altman analysis resulted in bias of 3.8% and limits of agreement of -13% and 21%.ConclusionsThe new mechanomyograph resulted in similar train-of-four ratio measurements compared to an archival mechanomyography system utilizing an FT-10 Grass force transducer. These results demonstrated continuity of gold standard measurement of neuromuscular blockade spanning nearly 50 yr, despite significant changes in the instrumentation technology.Copyright © 2023 American Society of Anesthesiologists. All Rights Reserved.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…