• Critical care medicine · Jan 1996

    Single breath CO2 analysis: description and validation of a method.

    • J H Arnold, J E Thompson, and L W Arnold.
    • Department of Anesthesia, Children's Hospital, Boston, MA 02115, USA.
    • Crit. Care Med. 1996 Jan 1;24(1):96-102.

    ObjectivesTo evaluate the performance of a newly developed single breath CO2 analysis station in measuring the airway deadspace in a lung model (study 1), and then to quantify the bias and precision of the physiologic deadspace measurement in a surfactant-depleted animal model (study 2).DesignA prospective bench validation of a new technique of airway deadspace measurement using a criterion standard (study 1); a prospective, animal cohort study comparing a new technique of physiologic deadspace measurement with a reference method (Bohr-Enghoff method) (study 2).SettingA bench laboratory and animal laboratory in a university-affiliated medical center.SubjectsA lung model (study 1), and adult sheep with induced surfactant deficiency (saline lavage) (study 2).MethodsThe single breath CO2 analysis station consists of a mainstream capnometer, a variable orifice pneumotachometer, a signal processor, and computer software with capability for both on- and off-line data analysis. Study 1: We evaluated the accuracy of the airway deadspace calculation using a plexiglass lung model. The capnometer and pneumotachometer were placed at the ventilator Y-piece with polyvinyl chloride tubing added to simulate increased airway deadspace. Segments of tubing were sequentially removed during each testing session to simulate decreasing deadspace. The calculated airway deadspace was derived from the single breath CO2 plot and compared with the actual tubing volume using least-squares linear regression and paired t-tests. Study 2: The accuracy of the physiologic deadspace measurement was examined in a saline-lavaged animal model by comparing the physiologic deadspace calculated from the single breath CO2 analysis station with values obtained using the Enghoff modification of the Bohr equation: deadspace/tidal volume ratio = (PaCO2-mixed expired PCO2)/PaCO2.Measurements And Main ResultsStudy 1: Thirty-six measurements of calculated airway deadspace were made and compared with actual circuit deadspace during four different testing conditions. Measured airway deadspace correlated significantly with actual circuit deadspace (r2 = .99). The proportional error of the method was -0.8% with a 95% confidence interval from -3.6% to 1.9%. Study 2: A total of 27 pairs of measurements in four different animals were available for analysis. The derived physiologic deadspace/tidal volume ratio significantly correlated with the value obtained using the Bohr-Enghoff method (r2 = .84). The bias and precision of our physiologic deadspace calculation were .02 and .02, respectively, and the mean percent difference for the physiologic deadspace calculated from the single breath CO2 analysis station was 2.4%.ConclusionsOur initial experience with the single breath CO2 analysis station indicates that this device can reliably provide on-line evaluation of the single-breath CO2 waveform. In particular, estimation of the airway and physiologic deadspace under a variety of testing conditions was consistently within 5% of actual values. We feel that with further application and refinement of the technique, single breath CO2 analysis may provide a noninvasive, on-line monitor of changes in pulmonary blood flow.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…