• Med. Sci. Monit. · Mar 2010

    Comparative Study Clinical Trial

    Comparison of arterial pressure cardiac output monitoring with transpulmonary thermodilution in septic patients.

    • Simon F Böettger, Dragan Pavlovic, Matthias Gründling, Michael Wendt, Orlando Hung, Dietrich Henzler, Hartmut Richard Kern, and Christian Lehmann.
    • Department of Anesthesiology and Intensive Care Medicine, Ernst-Moritz-Arndt-University Greifswald, Greifswald, Germany.
    • Med. Sci. Monit. 2010 Mar 1;16(3):PR1-7.

    BackgroundThe measurement of cardiac output in critically ill patients is complicated by rapid pathophysiological changes. The aim of this study was to compare the recently developed Arterial Pressure Cardiac Output algorithm (APCO) with transpulmonary thermodilution (TDCO). Clinical and hemodynamic parameters were tested for their impact on the measurements.Material/MethodsTwenty septic patients were examined. Cardiac output measurements were performed simultaneously on 3 consecutive days. The data were evaluated using regression analysis and the Bland Altman approach.ResultsBland Altman analysis presented a bias of 0.72 L/min and limits of agreement of 2.16 to 3.61 L/min for TDCO vs. APCO. Statistically significant covariables in the regression analysis were systemic vascular resistance (p<0.001), mean arterial pressure (p<0.001), cardiac function index (p=0.01), global end-diastolic index (p=0.02) and stroke volume index (p=0.005). Multiple linear regression analysis showed the residual percentage error decreased from 49.1% to 21.5%.ConclusionsThe APCO algorithm provides a broad range of hemodynamic measurements with a minimally invasive approach and simple access to the patient's hemodynamic state. However, an underestimation at high cardiac output and an overestimation at low cardiac output relative to transpulmonary thermodilution were observed in septic patients. Therefore, the APCO algorithm in its current state cannot be substituted for transpulmonary thermodilution.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.