J. Thorac. Cardiovasc. Surg. · Oct 2024
Transapical Ventricular Reshaping Reduces Functional Mitral Regurgitation and Improves Ventricular Function in a Preclinical Model of Ischemic Cardiomyopathy.
- Chase L King, Daisuke Onohara, Stephanie K Tom, Michael Silverman, Yuta Kikuchi, Brooks A Lane, Keawepono Wong, Julia R Toma, Amanda Maddamma, and Muralidhar Padala.
- Structural Heart Research and Innovation Laboratory, Cardiothoracic Research Laboratories, Carlyle Fraser Heart Center at Emory University Hospital Midtown, Atlanta, Ga; Division of Cardiothoracic Surgery, Emory University School of Medicine, Atlanta, Ga.
- J. Thorac. Cardiovasc. Surg. 2024 Oct 1; 168 (4): 10621071.e11062-1071.e1.
ObjectiveA significant proportion of patients with advanced heart failure present with dilated left ventricles and functional mitral regurgitation. These patients currently have limited treatment options. The MitraClip device (Abbott) has benefited only patients with smaller left ventricles (end-diastolic dimension <70 mm), whereas those with larger left ventricles did not benefit. A possible explanation is correcting functional mitral regurgitation alone may not adequately reduce the wall stresses of a dilated left ventricle. We have developed a beating-heart device that not only approximates the papillary muscles to reduce functional mitral regurgitation but also modifies the left ventricle size and shape to reduce wall stress.MethodsYorkshire swine (n = 16) had a myocardial infarction induced by permanent occlusion of the left circumflex with intracoronary ethanol. Three months later, the animals developed heart failure and moderate or greater functional mitral regurgitation. Through a transapical approach, the new device was implanted under echocardiography guidance to reshape the left ventricle and correct functional mitral regurgitation. The acute impact of this approach on the mitral valve and left ventricle was assessed with echocardiography and invasive hemodynamics.ResultsAfter reshaping, echocardiography showed a decrease in end-diastolic volume by 36.3 ± 30.5 mL (P < .001), a decrease in sphericity index by 0.143 ± 0.087 (P < .001), and an increase in ejection fraction of 5.90% ± 6.38% (P < .01). Mitral valve tenting area was reduced by 39.29 ± 33.66 mm2 (P < .001), coaptation length was increased by 2.12 ± 1.02 mm (P < .001), and posterior excursion angle was improved by 9.07° ± 9.14° (P < .01), resulting in functional mitral regurgitation reduction.ConclusionsCorrection of functional mitral regurgitation with favorable changes in mitral valve geometry and reduction in left ventricle geometry is possible with the proposed device.Copyright © 2023 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.