Comparative Study
Differences in glucose values obtained from point-of-care glucose meters and laboratory analysis in critically ill patients.
- Anjannette Cook, Delyla Laughlin, Margery Moore, Doreen North, Kathleen Wilkins, Gay Wong, Allyson Wallace-Scroggs, and Lisa Halvorsen.
- Providence Portland Medical Center, Portland, Oregon 97213, USA.
- Am. J. Crit. Care. 2009 Jan 1;18(1):65-71; quiz 72.
BackgroundBlood for glucose analysis is often obtained interchangeably from indwelling catheters and fingersticks.ObjectivesTo determine the level of agreement between glucose values obtained by laboratory analysis and with a point-of-care device for blood from 2 different sources: fingerstick and a central venous catheter.MethodsA method-comparison design was used. Point-of-care values for blood from fingersticks and catheters were compared with laboratory values for blood from catheters in a convenience sample of 67 critically ill patients. The effects of hematocrit level and finger edema on differences in glucose values between the 2 methods were also evaluated. A t test was used to determine differences in glucose values obtained via the 2 methods. Differences and limits of agreement were also calculated.ResultsLaboratory glucose values for blood from a catheter differed significantly from point-of-care values for blood from the catheter (t(1,66) = -9.18; P < .001) and from a fingerstick (t(1,66) = 6.53; P < .001). Glucose values for the 2 methods differed by 20 mg/dL or more for 1 of 6 patients (15%) for catheter samples and for 1 of 5 (21%) for fingerstick samples. Point-of-care glucose values for fingerstick and catheter samples did not differ (P = .98). Hematocrit level significantly explained the difference in glucose values between the 2 methods for both catheter (R(2) = 0.288; P < .001) and fingerstick (R(2) = 0.280; P = .02) samples.ConclusionsUse of a commonly used point-of-care device when precise glucose values are needed may lead to faulty treatment decisions.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.