• Respiratory care · Feb 2024

    Comparison of High-Frequency Oscillatory Ventilators.

    • Kaoru Okazaki and Jumpei Kuroda.
    • Drs Okazaki and Kuroda are affiliated with the Department of Neonatology, Tokyo Metropolitan Children's Medical Center, Tokyo, Japan. kaoru_okazaki@tmhp.jp.
    • Respir Care. 2024 Feb 28; 69 (3): 298305298-305.

    BackgroundThe performance of high-frequency oscillatory ventilators (HFOV) differs by the waveform generation mode and circuit characteristics. Few studies have described the performance of piston-type HFOV. The present study aimed to compare the amplitude required to reach the target high-frequency tidal volume ([Formula: see text]); determine the relationship between the settings and actual pressure in amplitude or mean airway pressure ([Formula: see text]); and describe the interaction among compliance, frequency, and endotracheal tube (ETT) inner diameter in 4 HFOV models, including Humming X, Vue (a piston type ventilator commonly used in Japan), VN500 (a diaphragm type), and SLE5000 (a reverse jet type).MethodsThe oscillatory ventilators were evaluated by using a 50-mL test lung with 0.5 and 1.0 mL/cm H2O compliance, [Formula: see text] of 10 cm H2O, frequency of 12 and 15 Hz, and ETT inner diameters 2.0, 2.5, and 3.5 mm. At each permutation of compliance, frequency, and ETT, the target high-frequency [Formula: see text] was increased from 0.5 to 3.0 mL. The change in [Formula: see text] from the ventilator (ventilator [Formula: see text]) to Y-piece (Y [Formula: see text]) and alveolar pressure (alveolar [Formula: see text]) and the change in amplitude from the ventilator (ventilator amplitude) to Y-piece (Y amplitude) and alveolar pressure (alveolar amplitude) were determined at high-frequency [Formula: see text] of 1.0 and 3.0 mL.ResultsTo achieve the target high-frequency [Formula: see text], the Humming X and Vue required a higher amplitude than did the SLE5000, but the maximum amplitude in the VN500 was unable to attain a larger high-frequency [Formula: see text]. Ventilator [Formula: see text] and alveolar pressure decreased at the Y-piece with the Humming X and Vue but increased with the SLE5000. The ventilator [Formula: see text] in the VN500 decreased remarkably at a frequency of 15 Hz. The ventilator amplitude in all 4 ventilators decreased while temporarily increasing at the Y-piece in the VN500.ConclusionsThe actual measured value, such as alveolar [Formula: see text] and high-frequency [Formula: see text], varied according to the type of HFOV system and the inner diameter of the ETT, even with identical settings. Clinicians should therefore determine the setting appropriate to each HFOV model.Copyright © 2024 by Daedalus Enterprises.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…