• Military medicine · Nov 2023

    Does Cuff Design Impact Insertion and Removal Force of Tracheostomy Tubes? A Bench Model.

    • Ulf Borg, Emmet Bolger, and Sean Morris.
    • Department of Medical Science, Respiratory Interventions, Medtronic, Boulder, CO 80301, USA.
    • Mil Med. 2023 Nov 8; 188 (Suppl 6): 629633629-633.

    IntroductionPercutaneous dilatational tracheostomy is a commonly performed procedure in intensive care units. Unrecognized tracheal ring fracture has been suggested as a possible factor for tracheal stenosis. The degree of tracheal compression relates to the amount of force required to cannulate the trachea. The objective of this study was to determine the force required to insert two types of tracheostomy tubes with different cuff designs.Materials And MethodsThis bench model measured the insertion and removal force of two tracheostomy tubes; one with a barrel-shaped, high-volume, low-pressure cuff (traditional Shiley tracheostomy tube) and another with a taper-shaped, low-volume, low-pressure cuff (Shiley flexible tracheostomy tube). Three sizes of tracheostomy tubes either with a barrel- or taper-shaped cuff were tested (Jackson sizes 4, 6, and 10, corresponding to 6.5-, 7.5-, and 10-mm ISO sizes, respectively). A model representing the tissue that the tube traverses to enter the tracheal lumen was designed, and the tracheostomy tube was mounted on a universal testing machine to measure the force necessary to insert and remove the tube.ResultsAcross all tracheostomy tubes' sizes tested, significantly less force was required to insert the Shiley flexible tracheostomy tube compared to the traditional Shiley tracheostomy tube. Significantly less force was also required to remove the flexible tracheostomy tube compared to the traditional tracheostomy tube.ConclusionsThis model suggests that less force is required to insert the Shiley flexible tracheostomy tube, which could result in less tracheal compression. This may be because of the smaller taper-shaped cuff that, when deflated, occupies less volume compared to the barrel-shaped cuff. As a result, less tracheal injury may occur when using the Shiley flexible tracheostomy tube during percutaneous tracheostomy procedures.© The Association of Military Surgeons of the United States 2023. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…