• Int J Clin Exp Med · Jan 2015

    An in vitro and in vivo validation of a novel color-coded syringe device for measuring the intracuff pressure in cuffed endotracheal tubes.

    • Mineto Kamata, Hiromi Kako, Archana S Ramesh, Senthil G Krishna, and Joseph D Tobias.
    • Department of Anesthesiology & Pain Medicine, Nationwide Children's Hospital Columbus, Ohio.
    • Int J Clin Exp Med. 2015 Jan 1;8(7):11356-9.

    BackgroundThe clinical practice of pediatric anesthesiology has changed with a transition to the use of cuffed endotracheal tubes (ETTs) in infants and children. The monitoring of intracuff pressure has been suggested as one means to limit the potential for damage to the tracheal mucosa. The current study evaluates the accuracy of a novel, color-coded syringe device which provides three zones (green, clear, and red) to estimate the intracuff pressure.MethodThe study was conducted in two phases. Phase 1 was an in vitro study where cuffed ETTs of sizes 4.0 mm, 5.0 mm and 6.0 mm ID were placed into polyvinylchloride tubing of appropriate sizes. A manometer and the syringe device were simultaneously attached to measure the intracuff pressure at the middle of the 3 different zones on the device (red, clear, and green). Phase 2 was an in vivo study where the syringe device and the manometer were simultaneously attached to the pilot balloon to measure the intracuff pressure and the corresponding zone on the color-coded syringe following endotracheal intubation. Statistical analysis included a descriptive reporting of the mean ± SD, median, range, and 95% confidence intervals (CI) of the actual intracuff pressure readings at the three zones of the syringe device during both its in vitro and in vivo use.ResultsFor phase 1 of the study, the 95% CI for the green, clear, and red zones were 21.5-21.8, 29.2-29.5, and 46.5-47.4 cmH2O respectively. This correlated well with the manufacturer reported values of 20-30, 30-40, and 40-60 cmH2O for the 3 zones (green, clear, and red respectively). Phase 2 of the study included 200 patients ranging in age from 0.1 to 21.8 years (6.7 ± 5.1 years) and in weight from 4.0 to 129.1 kilograms (29.4 ± 23.3 kgs). The size of the ETTs ranged from 3.0 to 7.0 mm ID. The intracuff pressure measured by the manometer ranged from 4 to 65 cmH2O (27.6 ± 9.7 cmH2O). The 95% CI for the green, clear, and red zones were 20.5-21.7, 27.7-29.1, and 41.2-46.5 cmH2O respectively. There was no significant differences noted when comparing different patient ages or sizes of ETT.ConclusionThe current study demonstrates a clinically acceptable correlation between the zones on this novel, color-coded syringe device and the actual measurement of the intracuff pressure obtained by a manometer for both in vitro and in vivo use. This device is a simple, reliable, portable and affordable method to monitor intracuff pressure.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…