• Spine · May 2024

    Lumbar Total Disc Replacement Device Removals and Revisions Performed During a 20 Year Experience with 2,141 Patients.

    • Richard D Guyer, Scott L Blumenthal, Jessica L Shellock, Jack E Zigler, and Donna D Ohnmeiss.
    • Center for Disc Replacement at Texas Back Institute; Plano, TX.
    • Spine. 2024 May 15; 49 (10): 671676671-676.

    Study DesignThis was a retrospective study with prospective patient contact attempted to collect current data.ObjectiveThe purpose was to investigate the incidence and reasons for lumbar total disk replacement (TDR) removal or revision.Summary Of Background DataA concern regarding lumbar TDR was safety, particularly the need for device removal or revision. This may be particularly important considering removal/revision requires repeat anterior exposure with an increased risk of vascular injury.MethodsData were collected for a series of 2141 lumbar TDR patients, beginning with the first case experience in 2000. The mean follow-up was 78.6 months. For each case of device removal/revision, the reason, duration from index surgery, and procedure performed were recorded.ResultsOf 2141 patients, 27 (1.26%) underwent TDR removal or revision. Device removal was performed in 24 patients (1.12%), while three patients underwent revision (0.14%). Of the 24 removals, 12 were due to migration and/or loosening, three developed problems post-trauma, two developed lymphocytic reaction to device materials, two had ongoing pain, and there was one case of each: TDR was too large, vertebral body fracture (osteoporosis), lytic lesion, device subsidence and facet arthrosis, and infection seeded from a chest infection 146 months post-TDR. The three revisions were for Core repositioning (technique error), device repositioning after displacement, and core replacement due to wear/failure. With respect to timing, 37.0% of removals/revisions occurred within one-month postimplantation. Of note, 40.7% of removals/revisions occurred in the first 25 TDR cases performed by individual surgeons. There was one significant vascular complication occurring in a patient whose TDR was removed due to trauma. This was also the only patient among 258 with ≥15-year follow-up who underwent removal/revision.ConclusionIn this large consecutive series, 1.26% of TDRs were removed/revised. The low rate over a 20 year period supports the safety of these devices.Copyright © 2024 Wolters Kluwer Health, Inc. All rights reserved.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…