Singapore medical journal · Feb 2024
Local cOinage and Hospital Equipment Index (Lo Hei): projectile distance of Singapore coinage and healthcare-related equipment in a 3T magnetic resonance imaging scanner.
- Shao Jin Ong, HallinanJames TpdJTDepartment of Diagnostic Imaging, National University Hospital, Singapore.Department of Diagnostic Radiology, Yong Loo Lin School of Medicine, National University of Singapore, Singapore., Deborah Khoo, Desmond Hoon, Koon Liang Chia, Joanne Hang, Lycia Teo, Peijing Su, Michael Ong, Bertrand Ang, and Swee Tian Quek.
- Department of Diagnostic Imaging, National University Hospital, Singapore.
- Singapore Med J. 2024 Feb 1; 65 (2): 616761-67.
IntroductionModern magnetic resonance imaging (MRI) scanners utilise superconducting magnets that are permanently active. Patients and healthcare professionals have been known to unintentionally introduce ferromagnetic objects into the scanning room. In this study, we evaluated the projectile risk of Singapore coinage as well as some common healthcare equipment within a 3 T MRI scanner.MethodsA rig termed 'Object eNtry Guidance and Linear Acceleration Instrument' (ONG LAI) was custom-built to facilitate safe trajectory of the putative ferromagnetic objects. A ballistic gel target was utilised as a human tissue surrogate to estimate tissue penetration. The point at which objects would self-propel towards the scanner was named 'Huge Unintended Acceleration Towards Actual Harm (HUAT AH)'.ResultsSingapore third-series coins (10-cent to 1-dollar coins) are highly ferromagnetic and would accelerate towards the MRI scanner from more than one metre away. Cannulas with their needles are ferromagnetic and would self-propel towards the scanner from a distance of 20 cm. Standard surgical masks are ferromagnetic and may lose their sealing efficacy when they are worn too close to the magnet. Among the tested objects, a can of pineapple drink (Lee Pineapple Juice) had the highest HUAT AH at a distance of more than 1.5 m.ConclusionSome local coinage and commonly found objects within a healthcare setting demonstrate ferromagnetic activity with projectile potential from a distance of more than 1 m. Patients and healthcare professionals should be cognisant of the risk associated with introducing these objects into the MRI scanning room.Copyright © 2024 Copyright: © 2024 Singapore Medical Journal.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.