Randomized Controlled Trial Comparative Study Clinical Trial
Comparison of ventilation performance: standard resuscitation bag and the resuscitation bag controller.
- T E Terndrup, R A Cherry, and J B McCabe.
- Department of Critical Care, State University of New York Health Science Center, Syracuse 13210.
- J Emerg Med. 1990 Mar 1;8(2):121-5.
AbstractClinical evaluation of ventilation performance during resuscitation is largely subjective. A mechanical device, the resuscitation bag controller (RC), which encircles the bag and allows controlled compression may improve the precision and accuracy of ventilation with manual resuscitation bags (MRB). We hypothesize that more precise, controlled pressure ventilation can be delivered with the RC, compared to the MRB. Prehospital (N = 13) and hospital personnel (N = 12) who routinely perform manual ventilation were randomized to either method of ventilation. Operators were instructed to ventilate an intubated adult mannequin. The percent of breaths delivered within a specified range, 800 to 1200 mL, was compared using simple regression analysis. The precision of tidal volume (TV) and peak airway pressure (PAP) was compared between methods and groups using the coefficient of variation. Comparison for significant differences between methods and groups in the number of breaths with a TV less than 800 mL and those with pressures greater than 30 cm H20 was performed using chi square or Fisher's exact test. There were no significant differences in the percent of acceptable breaths or mean TV delivered between methods or groups. The precision (i.e., reproducibility of the same value) of TV for both methods was best for hospital personnel. PAPS were less than or equal to 30 cm H2O for 93% of all breaths. Hospital personnel delivered a significantly greater percent of inadequate (less than 0.8 L) breaths, 19 versus 7.4%, and excessive pressure breaths, 9.2 versus 4.2%, when compared to prehospital personnel. We conclude that the resuscitation bag controller offers little advantage over standard bag resuscitation for adult resuscitation.(ABSTRACT TRUNCATED AT 250 WORDS)
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.