• Health devices · Oct 2010

    User experience network. Erroneous downstream occlusion alarms may disable Smiths Medical CADD-Solis infusion pumps.

    • Health Devices. 2010 Oct 1;39(10):380-1.

    AbstractDue to an issue in manufacturing, downstream occlusion (DSO) sensors in some Smiths Medical CADD-Solis infusion pumps may drift out of calibration, potentially resulting in erroneous alarms that disable the units. Hospitals experiencing the problem should return affected units to Smiths Medical for recalibration (free of charge) and should consider testing all their CADD-Solis pumps during routine maintenance to ensure that they alarm appropriately for downstream occlusions.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…