• J. Thorac. Cardiovasc. Surg. · Dec 2010

    Reproducibility of left atrial ablation with high-intensity focused ultrasound energy in a calf model.

    • Nestor R Villamizar, Jennifer H Crow, Valentino Piacentino, Louis R DiBernardo, Mani A Daneshmand, Dawn E Bowles, Mark A Groh, and Carmelo A Milano.
    • Department of Surgery, Duke University Medical Center, Durham, NC 27713, USA. villa022@mc.duke.edu
    • J. Thorac. Cardiovasc. Surg. 2010 Dec 1; 140 (6): 13817.e11381-7.e1.

    ObjectiveAchieving transmural tissue ablation might be necessary for successful treatment of atrial fibrillation. The purpose of this study was to evaluate the reproducibility of transmural left atrial ablation using a high-intensity focused ultrasound energy system in a calf model.MethodsNine heparinized bovines underwent a beating-heart left atrial ablation with a single application of the high-intensity focused ultrasound device. All animals were acutely killed, and the left atrium was fixed in formalin. Protocolized histological sections (5 μm) were obtained throughout each lesion and prepared with Masson trichrome and hematoxylin and eosin staining. Measurements were performed on a total of 359 slides from the 9 lesions. In addition, fresh left atrial tissues from 18 unused human donor hearts that did not meet the criteria for cardiac transplantation were measured at the site where the high-intensity focused ultrasound device is normally applied.ResultsCalf left atrial thickness ranged between 2.5 and 20.1 mm, with a mean of 9.10 mm. High-intensity focused ultrasound ablation consistently produced a 100% transmural lesion in left atrial thickness up to 6 mm. In addition, a transmural lesion was observed in 91% of tissues that were up to 10 mm thick and in 85% that were up to 15 mm thick. Human left atrial thickness ranged between 1.2 to 6 mm, with a mean of 3.7 mm.ConclusionsCalf left atrial thickness in this study was greater than human left atrial thickness. Human left atrial thickness is generally less than 6 mm, and in this range high-intensity focused ultrasound ablation achieved 100% transmurality. These histological results might correlate with a high success rate of atrial fibrillation ablation by using the high-intensity focused ultrasound system.Copyright © 2010 The American Association for Thoracic Surgery. Published by Mosby, Inc. All rights reserved.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…