• Eur Spine J · Jul 2024

    Endplate weakening during cage bed preparation significantly reduces endplate load capacity.

    • Anna-Katharina Calek, Frédéric Cornaz, Mauro Suter, Marie-Rosa Fasser, Mazda Farshad, and Jonas Widmer.
    • Department of Orthopedics, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland. anna-katharina.calek@balgrist.ch.
    • Eur Spine J. 2024 Jul 1; 33 (7): 262126292621-2629.

    PurposeTo analyze the effect of endplate weakness prior to PLIF or TLIF cage implantation and compare it to the opposite intact endplate of the same vertebral body. In addition, the influence of bone quality on endplate resistance was investigated.MethodsTwenty-two human lumbar vertebrae were tested in a ramp-to-failure test. One endplate of each vertebral body was tested intact and the other after weakening with a rasp (over an area of 200 mm2). Either a TLIF or PLIF cage was then placed and the compression load was applied across the cage until failure of the endplate. Failure was defined as the first local maximum of the force measurement. Bone quality was assessed by determining the Hounsfield units (HU) on CT images.ResultsWith an intact endplate and a TLIF cage, the median force to failure was 1276.3N (693.1-1980.6N). Endplate weakening reduced axial endplate resistance to failure by 15% (0-23%). With an intact endplate and a PLIF cage, the median force to failure was 1057.2N (701.2-1735.5N). Endplate weakening reduced axial endplate resistance to failure by 36.6% (7-47.9%). Bone quality correlated linearly with the force at which endplate failure occurred. Intact and weakened endplates showed a strong positive correlation: intact-TLIF: r = 0.964, slope of the regression line (slope) = 11.8, p < 0.001; intact-PLIF: r = 0.909, slope = 11.2, p = 5.5E-05; weakened-TLIF: r = 0.973, slope = 12.5, p < 0.001; weakened-PLIF: r = 0.836, slope = 6, p = 0.003.ConclusionWeakening of the endplate during cage bed preparation significantly reduces the resistance of the endplate to subsidence to failure: endplate load capacity is reduced by 15% with TLIF and 37% with PLIF. Bone quality correlates with the force at which endplate failure occurs.© 2024. The Author(s).

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,624,503 articles already indexed!

We guarantee your privacy. Your email address will not be shared.