• Terapevt Arkh · Dec 2023

    [Point-of-Care Blood Glucose Testing: Post-Market Performance Assessment of the Accu-Chek Inform II Hospital-Use Glucose Meter].

    • E E Petryaykina, N A Mayanskiy, E S Demina, I V Karamysheva, K A Gorst, and A V Timofeev.
    • Russian Children's Clinical Hospital - Branch of Pirogov Russian National Research Medical University.
    • Terapevt Arkh. 2023 Dec 28; 95 (12): 115111631151-1163.

    BackgroundA point-of-care glucose testing (POCT) is an essential component of care in patients with hyperglycemia and hypoglycemia in inpatient and outpatient settings. In Russian medical facilities (MFs), conventional glucose meters designed for self-monitoring by patients with diabetes are commonly used for POCT. These home-use meters have two serious disadvantages: the first is large measurement bias and the second - they can't be integrated into laboratory information systems, so measurement data have to be recorded into patient charts manually. Both factors may lead to medical errors. It is reasonable to use in the MFs specialized POCT glucose meters, as they are superior to conventional ones in accuracy and may be easily connected to laboratory information systems. With this in mind, physicians at the Russian Children's Clinical Hospital decided to substitute conventional meters with the Accu-Chek Inform II POCT meter, however, after preliminary performance assessment of the model.AimTo test the Accu-Chek Inform II performance characteristics: accuracy, linearity, repeatability, and mean absolute relative difference (MARD).Materials And MethodsPerformance of the Accu-Chek Inform II was tested by comparing the results of parallel CGL measurements with the meter and reference laboratory analyzer in capillary blood samples. Overall, 99 parallel CGL measurements were made in 45 samples. Accuracy was evaluated according to the ISO 15197-2013 and POCT12-A3 criteria.ResultsThe Accu-Chek Inform II meter met the requirements of ISO 15197-2013 and POCT12-A3 and demonstrated high linearity (correlation coefficient, r=1,0), good repeatability (mean coefficient of variation, CV=1,38%) and acceptable MARD (4,9%).ConclusionThe Accu-Chek Inform II POCT glucose meter may be efficiently and safely used in inpatient and outpatient MFs and particularly in pediatric clinics.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…