Arch Orthop Trauma Surg · Aug 2024
Primary stability of multi-hole acetabular cup combined with posterior column plating for the fixation of complex acetabular fractures in elderly patients: a biomechanical analysis.
- Shuang G Yan, Zhi Wang, Yiliang Cui, Xingyi Hua, Fanxiao Liu, and Di Li.
- Department of Orthopaedic Surgery, The First Affiliated Hospital of Anhui Medical University, Hefei, China. yanshuanggen686@hotmail.com.
- Arch Orthop Trauma Surg. 2024 Aug 1; 144 (8): 369737033697-3703.
BackgroundAcute total hip arthroplasty has gained increasing popularity in the treatment of complex acetabular fractures in elderly patients, but the biomechanical evidence is lacking. Therefore, we measured the primary stability to evaluate whether acute total hip arthroplasty using a multi-hole acetabular cup and posterior column plating through a single K-L approach can safely be used for treating complexed acetabular fractures in the elderly.MethodsIn 18 composite osteoporotic hemipelves, T-type acetabular fractures were treated in three ways: CSP (Cup with acetabular screws augmentation combined with posterior plate) group, CP (Cup without acetabular screw combined with posterior plate) group and CSPA (Cup with acetabular screws combined with posterior plate and anterior plate) group. Each specimen was dynamically loaded (300 to 1700 N, 1 Hz). Primary stability of the acetabular cup and fracture gap was evaluated by 3-dimensional (3D) micromotions.ResultsNo significant differences in the 3D-micromotions were observed among the CSP, CP and CSPA groups except the test point on the anterior column of the acetabulum in the CP scenario (CSP 49.33 ± 21.08 μm vs. CP 224.83 ± 52.29 μm, p < 0.001; CSPA 45.50 ± 12.16 μm vs. CP 224.83 ± 52.29 μm, p < 0.001). No significant differences in the fracture gap displacement on the posterior column of acetabulum were observed among the CSP, CP and CSPA groups.ConclusionOur results show, that acute total hip arthroplasty using a multi-hole cup with acetabular screws and posterior column plating through single a K-L approach offers good primary stability to allow good osseous integration for treating complex acetabular fractures in the elderly. Furthermore, it also provides good fracture gap displacement on the posterior column.© 2024. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.