Comparative Study
Biomechanical analysis of various internal fracture fixation devices used for treating femoral neck fractures: A comparative finite element analysis.
- Biswajit Mahapatra and Bidyut Pal.
- Department of Mechanical Engineering, Indian Institute of Engineering Science and Technology, Shibpur, Howrah 711103, West Bengal, India.
- Injury. 2024 Oct 1; 55 (10): 111717111717.
IntroductionSeveral internal fixation devices are available for treating Pauwels type I, II and III femoral neck fractures. The present study compared various fixation implants for all Pauwels fracture types using a CT-based subject-specific finite element model of the femur and determined the most effective implant for each fracture type.Materials And MethodsThe analysis included four different configurations of cannulated screw models, Femoral Neck System, Dynamic Hip Screw and Dynamic Condylar Screw (with and without anti-rotational screw). Ti-alloy was considered as the implant material. Heterogeneous bone material property was assigned based on CT grey value. Frictional contact was assumed in the contact interfaces. Peak loading corresponding to normal walking and stair-climbing were considered. Equivalent strain in bone, equivalent stress in the implants, femoral head deformation and rotation, micromotion in the contact interfaces, and strain-shielding in bone were evaluated for each implanted model.ResultsStresses generated in the implants were within the yield limit of the implant material. In Pauwels I and II, the micromotion predicted at the contact regions in all the implanted models was within 100 µm, which is suitable for bone integration. However, in Pauwels III fracture, most of the implanted models other than DHS with AR-screw model exhibited micromotion of more than 150 µm in the contact regions, which is expected to inhibit bone growth.ConclusionsThe DHS with AR-screw implanted model was identified as the most effective in treating Pauwels I and III fractures. However, for Pauwels type II, DCS with an AR-screw implant was deemed superior to the other configurations.Copyright © 2024 Elsevier Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.