    In vitro evaluation of flow diverter performance using a human fibrinogen-based flow model.

    • Cem Bilgin, Esref Alperen Bayraktar, Alexander A Oliver, Jiahui Li, Juan R Cebral, David F Kallmes, and Ramanathan Kadirvel.
    • Departments of1Radiology and.
    • J. Neurosurg. 2024 Jul 12: 181-8.

    ObjectiveFibrin deposition represents a key step in aneurysm occlusion, promoting endothelization of implants and connective tissue organization as part of the aneurysm-healing mechanism. In this study, the authors introduce a novel in vitro testing platform for flow diverters based on human fibrinogen.MethodsA flow diverter was deployed in 4 different glass models. The glass models had the same internal parent artery (4 mm) and aneurysm (8 mm) diameters with varying parent artery angulations (paraophthalmic, sidewall, bifurcation, and slightly curved models). The neck size and area were 4 mm and 25 mm2, respectively. Human fibrinogen (330 mg/dl) was circulated within the glass models at varying flow rates (0, 3, 4, and 5 ml/sec) with or without heparin, calcium chloride, and thrombin for as long as 6 hours or until complete fibrin coverage of the flow diverter's neck was achieved. Aneurysm neck coverage was defined as macroscopic fibrin deposition occluding the flow diverters' pores. Flow characteristics after flow diverter deployment were assessed with computational fluid dynamics analysis. The effects of flow rates, heparin, calcium chloride, and thrombin on fibrin deposition rates were tested using 1-way ANOVA and the Tukey test.ResultsA total of 84 replicates were performed. Human fibrin did not accumulate on the flow diverter stents under static conditions. The fibrin deposition rate on the aneurysm neck was significantly greater with the 5 ml/sec flow rate as compared to 3 ml/sec for all models. The paraophthalmic model had the highest inflow velocity of 48.7 cm/sec. The bifurcation model had the highest maximum shear stress (SS) and maximum normalized shear stress values at the device cells at 843.3 dyne/cm2 and 35.1 SS/SSinflow, respectively. The fibrin deposition rates of the paraophthalmic and bifurcation models were significantly higher than those of sidewall and slightly curved models for all additive or flow rate comparisons (p = 0.001 for all comparisons). The incorporation of thrombin significantly increased the fibrin deposition rates across all models (p = 0.001 for all models).ConclusionsRates of fibrin deposition varied widely across different configurations and additive conditions in this novel in vitro model system. Fibrin accumulation started at the aneurysm inflow zone where flow velocity and shear stress were the highest. The primary factors influencing fibrin deposition included flow velocities, shear stress, and the addition of thrombin at a physiological concentration. Further research is needed to test the clinical utility of fibrinogen-based models for patient-specific aneurysms.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…