• Military medicine · Aug 2024

    Evaluation of Inhaled Nitric Oxide Generation Systems at Altitude.

    • Thomas Blakeman, Dario Rodriquez, Maia Smith, Michael Goodman, and Richard Branson.
    • Department of Surgery, University of Cincinnati School of Medicine, Cincinnati, OH 45267, USA.
    • Mil Med. 2024 Aug 19; 189 (Suppl 3): 579584579-584.

    IntroductionInhaled nitric oxide (INO) is a selective pulmonary vasodilator delivered from compressed gas cylinders filled to 2,200 psig (137.8 bar) with 800 ppm of NO in a balance of nitrogen. NO is currently FDA-approved for use in term or near-term infants with hypoxemia and signs of pulmonary hypertension in the absence of cardiac disease. INO has also been shown to improve oxygenation in adults with refractory hypoxemia. Current doctrine precludes the use of NO during military aeromedical transport owing to the requirement for large compressed gas cylinders. We performed a bench evaluation of 2 delivery systems that create NO from room air without the need for pressurized cylinders.Materials And MethodsWe evaluated 2 portable nitric oxide INO generation systems (LungFit PH, Beyond Air Inc, Garden City, NJ and a prototype NO generator, Odic Inc, Littleton, MA) at ground level, 8,000, and 14,000 feet (2,437 and 4,267 meter) simulated altitude in an altitude chamber. The output from each device was injected into the inspiratory limb of the ventilator circuit that was attached to a test lung. A 731 ventilator (Zoll Medical, Chelmsford, MA) and T1 (Hamilton Medical, Reno, NV) were used employing 24 combinations of ventilator settings each repeated in duplicate. An INOmax DS IR was used to measure delivered INO and NO2 via a sampling line attached in the ventilator circuit inspiratory limb. A fast response oxygen analyzer (O2CAP, Oxigraf Inc, Sunnyvale, CA) was used to measure inspired FiO2. Target INO concentration was 20 ppm.ResultsAcross all ventilator settings, the LungFit device delivered INO was 19.8 ± 1.6 ppm, 16.1 ± 1.9 ppm, and 11.6 ± 1.7 ppm at ground level, 8,000 ft (2,437 meter), and 14,000 ft (4,267 meter), respectively. The Odic device delivered INO dose was 20.6 ± 1.4 ppm, 21.3 ± 5.5 ppm, and 20.4 ± 9.1 ppm at ground level, 8,000 ft (2,437 meter), and 14,000 ft (4,267 meter), respectively.ConclusionsBoth devices delivered a reliable INO dose at ground level. Altitude significantly affected INO delivery accuracy at 14,000 ft (4,267 meter) (P < 0.01) with both devices and at 8,000 ft (2,437 meter) (P < 0.01) with LungFit. Differences in INO dosage were not statistically significant with the Odic device at 8,000 ft (2,437 meter)(P > 0.05) although there were large variations with selected ventilator settings. With careful monitoring, devices creating INO from room air without cylinders could be used during aeromedical transport without the need for pressurized cylinders.© The Association of Military Surgeons of the United States 2024. All rights reserved. For commercial re-use, please contact reprints@oup.com for reprints and translation rights for reprints. All other permissions can be obtained through our RightsLink service via the Permissions link on the article page on our site–for further information please contact journals.permissions@oup.com.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…