• Neurocritical care · Sep 2024

    Noninvasive Method Using Mechanical Extensometer for the Estimation of Intracranial Compliance by Repeated Measures Agreement Analysis.

    • Sanem Pinar Uysal, Hayley G Williams, Mina Huerta, Nicolas R Thompson, and Catherine E Hassett.
    • Department of Neurology, Neurological Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH, 44195, USA. sanempinaruysal@gmail.com.
    • Neurocrit Care. 2024 Sep 18.

    BackgroundIntracranial compliance refers to the relationship between changes in volume and the resultant changes in intracranial pressure (ICP). This study aimed to assess the agreement of a noninvasive ICP waveform device for the estimation of compliance compared with invasive ICP monitoring employing three distinct methods.MethodsWe conducted a retrospective analysis of ICP waveform morphology recorded through both invasive (external ventricular drain) and noninvasive (mechanical extensometer) methods in adult patients with acute brain injury admitted to a neurointensive care unit between August 2021 and August 2022. Compliance was calculated as the amplitude of the fundamental component of cerebral arterial blood volume (estimated with concurrent Transcranial Doppler [TCD] recordings), divided by the amplitude of the fundamental component of the invasive and noninvasive ICP waveforms. Subsequently, we assessed the agreement between invasive and noninvasive intracranial compliance by repeated measures correlation coefficient analysis using three methods: TCD-derived, P2/P1 ratio, and time-to-peak (TTP). A linear mixed-effects model was used to compute the concordance correlation coefficient, total deviation index, and coefficient of individual agreement. Coverage probability plot was calculated to estimate the percent of observations within different cut points for each of the three methods.ResultsA total of 21 patients were identified for this study. Repeated measures correlation analysis showed a strong correlation (R = 0.982, 95% confidence interval [0.980-0.984], p < 0.0001) between log-transformed noninvasive and invasive compliance. Agreement statistics for TCD, P2/P1 ratio, and TTP indicated that although the concordance correlation coefficient was highest for log(TCD) values, TTP and P2:P1 ratio measures had better agreement with total deviation index and coverage probability plot analyses.ConclusionsRepeated measures correlations suggest that ICP waveform analyses may offer a more accurate estimate of compliance than TCD-derived methods for noninvasive ICP monitoring. Further validations studies are warranted to confidently establish this method as an indicator of intracranial compliance.© 2024. Springer Science+Business Media, LLC, part of Springer Nature and Neurocritical Care Society.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…