• Can J Anaesth · Sep 2024

    Comparison of the ClearSight™ finger cuff monitor versus invasive arterial blood pressure measurement in elective cardiac surgery patients: a prospective observational study.

    • Una Goncin, Kaixuan K Liu, Brooklyn Rawlyk, Sara Dalkilic, Mary Ellen J Walker, Jonathan Norton, and Peter Hedlin.
    • Provincial Department of Anesthesiology, College of Medicine, University of Saskatchewan, Saskatoon, SK, Canada.
    • Can J Anaesth. 2024 Sep 24.

    PurposeTo determine the acceptability of the ClearSight™ system (Edwards Lifesciences Corp., Irvine, CA, USA) for continuous blood pressure monitoring during elective cardiac surgery compared with arterial catheterization.MethodsWe enrolled 30 patients undergoing elective cardiac surgery in a prospective observational study. Blood pressure measurements were recorded every 10 sec intraoperatively. We determined agreement based on the Association for the Advancement of Medical Instrumentation (AAMI) recommendations. Statistical analysis included fixed bias (difference of measurements between methods), percentage error (accuracy between ClearSight measurement and expected measurement from arterial line), and interchangeability (ability to substitute ClearSight monitor without effecting overall outcome of analysis). We used a paired samples t test to compare the time required for placing each monitor.ResultsWe found fixed bias in the differences between the ClearSight monitor and invasive arterial blood pressure measurement in systolic blood pressure (SBP; mean difference, 8.7; P < 0.001) and diastolic blood pressure (DBP; mean difference, -2.2; P < 0.001), but not in mean arterial pressure (MAP; mean difference, -0.5; P < 0.001). Bland-Altman plots showed that the means of the limits of agreement were greater than 5 mm Hg for SBP, DBP, and MAP. The percentage errors for SBP, DBP, and MAP were lower than the cutoff we calculated from the invasive arterial blood pressure measurements. Average interchangeability rates were 38% for SBP, 50% for DBP, and 50% for MAP. Placement of the ClearSight finger cuff was significantly faster compared with arterial catheterization (mean [standard deviation], 1.7 [0.6] min vs 5.6 [4.1] min; P < 0.001).ConclusionsIn this prospective observational study, we did not find the ClearSight system to be an acceptable substitute for invasive arterial blood pressure measurement in elective cardiac surgery patients according to AAMI guidelines. Nevertheless, based on statistical standards, there is evidence to suggest otherwise.Study RegistrationClinicalTrials.gov ( NCT05825937 ); first submitted 11 April 2023.© 2024. Canadian Anesthesiologists' Society.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…