The lancet oncology · Nov 2024
Xpert Breast Cancer STRAT4 Assay using fine-needle aspiration biopsy samples in a resource-constrained setting: a prospective diagnostic accuracy study.
- Dianna L Ng, Edda Vuhahula, Asteria H Kimambo, Marie Claire Ndayisaba, Godfrey S Philipo, Beatrice P Mushi, Kenneth E Ho, Alan Paciorek, Zainab Illonga, Li Zhang, Poonam Vohra, Jodi Weidler, Michael Bates, Elia J Mmbaga, and Katherine Van Loon.
- Department of Pathology, University of California, San Francisco, San Francisco, CA, USA; Helen Diller Family Comprehensive Cancer Center, University of California, San Francisco, San Francisco, CA, USA; Department of Pathology and Laboratory Medicine, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
- Lancet Oncol. 2024 Nov 1; 25 (11): 144014521440-1452.
BackgroundUse of fine-needle aspiration biopsy (FNAB) specimens on Xpert Breast Cancer STRAT4 Assay (STRAT4; Cepheid, Sunnyvale, CA, USA), a CE-marked in-vitro diagnostic medical device, could potentially increase access to breast cancer biomarker testing in resource-constrained settings. We aimed to assess the performance of a research use-only version of STRAT4 using FNAB specimens in Tanzania.MethodsIn this prospective diagnostic accuracy study, patients aged 18 years or older with palpable breast masses presenting to the FNAB Clinic at Muhimbili National Hospital (Dar es Salaam, Tanzania) were recruited consecutively. Patients who were pregnant, lactating, or had a previous diagnosis of breast cancer were excluded. STRAT4 testing was performed on off-label FNAB samples using four protocols: the 1 × protocol (using the standard lysate method) on FNAB smears (1 × FNAB), quick lysis and Maui protocols (both on FNAB smears), and the 1 × protocol on formalin-fixed paraffin-embedded (FFPE) cell block material (1 × cell block). For 1 × FNAB and 1 × cell block, tissue was processed using FFPE lysis reagent, incubated at 80°C with proteinase K, and followed by addition of 95% or higher ethanol. Quick lysis was processed using FFPE lysis reagent and 95% or higher ethanol, whereas Maui was processed using a proprietary research-use only lysis reagent. The primary outcomes were overall concordance, sensitivity, specificity, and area under the receiver operating characteristic curve (AUC) of STRAT4 as compared with immunohistochemistry or immunohistochemistry plus fluorescence in-situ hybridisation performed on cell blocks using clinically validated protocols in a Clinical Laboratory Improvement Amendments-accredited laboratory at the University of California, San Francisco (San Francisco, CA, USA).FindingsBetween Nov 29, 2017, and Dec 17, 2020, 208 patients were enrolled. Of 208 cases, 51 (25%) were excluded from analysis because of insufficient tissue in the cell block or absent cell blocks, leaving 157 participants (all female) for analysis. For oestrogen receptor, 1 × FNAB had the best performance, with an overall concordance of 95% (95% CI 90-100), sensitivity of 94% (85-100), specificity of 97% (90-100), and AUC of 0·96 (0·81-1·00). For progesterone receptor, 1 × cell block had the best overall performance (overall concordance 89% [95% CI 84-95], sensitivity 91% [82-99], and specificity 89% [81-97], with an AUC of 0·93 [0·89-0·99]) and 1 × FNAB performed the best among the smear protocols, with a concordance of 84% (95% CI 74-93), sensitivity of 63% (43-82), specificity of 97% (92-100), and AUC of 0·91 (0·72-0·97). For HER2, Maui had the highest agreement, with an overall concordance of 93% (95% CI 89-98), sensitivity of 96% (88-100), specificity of 92% (87-98), and AUC of 0·95 (0·98-1·00). For Ki67, Maui had the best performance of smear protocols, with a concordance of 73% (95% CI 64-82), sensitivity of 70% (58-81), specificity of 81% (66-96), and AUC of 0·80 (0·54-0·82).InterpretationProcessing FNAB samples with STRAT4 is feasible in Tanzania, and performance for the oestrogen receptor is robust. Further optimisation of STRAT4 for FNAB has the potential to improve timely access to breast cancer diagnostics in resource-constrained settings.FundingUS National Institutes of Health; UCSF Global Cancer Program, Helen Diller Family Comprehensive Cancer Center; UCSF Department of Pathology; and Cepheid.Copyright © 2024 Elsevier Ltd. All rights reserved, including those for text and data mining, AI training, and similar technologies.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.