• Anesthesia and analgesia · Dec 2008

    Preparation of the Dräger Fabius anesthesia machine for the malignant-hyperthermia susceptible patient.

    • Joel B Gunter, John Ball, and Sean Than-Win.
    • Department of Anesthesia, Children's Hospital Medical Center, 3333 Burnet Ave., Cincinnati, OH 45229, USA. joel.gunter@cchmc.org
    • Anesth. Analg. 2008 Dec 1;107(6):1936-45.

    BackgroundAnesthesia machines must be flushed of halogenated anesthetics before use in patients susceptible to malignant hyperthermia. We studied the kinetics of sevoflurane clearance in the Dräger Fabius anesthesia machine and compared them to a conventional anesthesia machine (Dräger Narkomed GS).MethodsBefore each trial, the anesthesia machine was contaminated for 2 h with 3% sevoflurane and then prepared by changing the CO(2) absorbent, removing the vaporizer(s), and mounting a clean circuit and artificial lung. The basic flush procedure consisted of oxygen 10 L/min with the ventilator set to a tidal volume of 600 mL at a rate of 10/min. Residual sevoflurane in the inspiratory limb of the circuit was measured using an ambient air analyzer capable of measuring sevoflurane to < 1 ppm. Results were analyzed using log-linear regression of residual concentration as a function of time. This model was used to estimate the time required to achieve a desired residual anesthetic concentration.ResultsTimes to achieve 10 and 5 ppm in the Dräger Narkomed GS were 11 and 18 min, respectively. For the Dräger Fabius anesthesia machine, times to 10 and 5 ppm were 75 and 104 min, respectively. Several maneuvers to accelerate clearance of residual sevoflurane from the Dräger Fabius resulted in only modest reductions in these times (10 and 5 ppm times 40-50 min and 60-80 min, respectively). Insertion of an activated charcoal filter (QED, Anecare Laboratories, Salt Lake City, UT) into the inspiratory limb of the Dräger Fabius circuit reduced the residual anesthetic concentration to <5 ppm within 10 min; this concentration was maintained for > 6 h despite a fresh gas flow of only 2 L/min after the first 15 min.DiscussionPreparation of the Dräger Fabius anesthesia machine using conventional flushing techniques required almost 10 times as long as an older, conventional anesthesia machine. If a prolonged flush is impractical or impossible, we describe a procedure using an activated charcoal filter inserted on the inspiratory limb of the breathing circuit which can effectively scrub residual sevoflurane to a concentration < 5 ppm within 10 min. This procedure includes an initial 5 min flush without the activated charcoal filter followed by a 5 min flush with the charcoal filter, after which the machine is ready for use in the malignant hyperthermia-susceptible patient. The charcoal filter must remain on the machine for the remainder of the anesthetic, and the fresh gas flow should be maintained > or = 10 L/min for the first 5 min, and > or = 2 L/min thereafter.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…